FDA Adverse Event Injury Summary report: N

MENTOR MEMORYSHAPE BREAST IMPLANT

MDR report key: 23122550 · Received September 23, 2025

Report

Report Number
1645337-2025-09935
Event Type
Injury
Date Received
September 23, 2025
Date of Event
August 13, 2025
Manufacturer
MENTOR TEXAS
Product Code
FTR
PMA / PMN Number
P060028
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON OCTOBER 01, 2025, MENTOR RECEIVED ADDITIONAL INFORMATION REGARDING THE PATIENT'S REPLACEMENT DEVICES. THE PATIENT'S REPLACEMENT DEVICES WERE TWO 370CC MENTOR MEMORYGEL BOOST BREAST IMPLANTS (CATALOG NUMBER SHPB370) WITH LOT NUMBER 2070778 AND 2062968. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON NOVEMBER 06, 2025, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. ON NOVEMBER 10, 2025, THE EVALUATION FOR THE DEVICE WAS COMPLETED. DEVICE EVALUATION SUMMARY: VISUAL ANALYSIS OF THE RETURNED DEVICE REVEALED THAT NO DAMAGE OR ANOMALIES WERE OBSERVED ON THE BREAST IMPLANT. LEAK TESTING WAS PERFORMED, IN ACCORDANCE WITH MENTOR PROCEDURES, AND NO AREAS OF GEL EXPOSURE WERE DETECTED DURING THE ANALYSIS. ALTHOUGH NO CONCLUSION COULD BE REACHED ON THE CAUSE OF THE REPORTED EVENT, THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING CAUTION: RUPTURE STATUS IDENTIFIED BY MRI EVALUATION INCLUDES BOTH SUSPECTED RUPTURES, WHICH ARE THOSE RUPTURES IDENTIFIED BY MRI BUT NOT CONFIRMED BY EXPLANTATION AND EXAMINATION OF THE DEVICE, AND CONFIRMED RUPTURES, WHICH ARE THOSE RUPTURES THAT ARE CONFIRMED BY EVALUATION OF THE EXPLANTED DEVICES. THE EVENT DESCRIBED COULD NOT BE CONFIRMED AS THE BREAST IMPLANT WAS RETURNED WITHOUT DETECTABLE GEL EXPOSURE. ALTHOUGH NO PRODUCT DEFECT WAS IDENTIFIED, THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT COULD NOT BE REPLICATED DURING THE LABORATORY ANALYSIS. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA: SINCE NO MALFUNCTION WAS OBSERVED DURING THE INVESTIGATION. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RUPTURE. D6B. EXPLANTATION DATE: (B)(6) 2025. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A 59-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH TWO 685CC MENTOR MEMORYSHAPE BREAST IMPLANTS AND EXPERIENCED BILATERAL RUPTURES POST-OPERATIVELY, WHICH WAS DIAGNOSED WITH A PHYSICAL EXAM. AS A RESULT, THE PATIENT IS TO UNDERGO DEVICE EXPLANTATION ON (B)(6) 2025. THIS REPORT IS FOR THE RIGHT-SIDE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
580349 MENTOR MEMORYSHAPE BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 7308738

Patients

Seq Age Sex Outcome Treatment
1 59 YR Female Required Intervention