VISTASEAL LAPAROSCOPIC DUAL APPLICATOR (45 CM FLEXIBLE)
Report
- Report Number
- 2210968-2025-10844
- Event Type
- Malfunction
- Date Received
- September 23, 2025
- Date of Event
- September 19, 2025
- Report Date
- November 20, 2025
- Manufacturer
- ETHICON INC.
- Product Code
- MZM
- UDI-DI
- 10705031462724
- PMA / PMN Number
- BK190324
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
PRODUCT COMPLAINT #: (B)(4). CORRECTED INFORMATION: D 9. DEVICE AVAILABLE FOR EVALUATION? ADDITIONAL INFORMATION: D 9. DATE DEVICE RETURNED TO MANUFACTURER, D 9. IS DEVICE RETURNED TO MANUFACTURER? H 6. COMPONENT CODE, H 6. TYPE OF INVESTIGATION, H 6. INVESTIGATION FINDINGS, H 6. INVESTIGATION CONCLUSIONS H6 COMPONENT CODE: G07002 - PENDING EVALUATION OF RETURNED DEVICE. THE DEVICE UPON WHICH THIS MEDWATCH IS BASED HAS BEEN RECEIVED, HOWEVER, THE EVALUATION IS NOT YET COMPLETE. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT#: (B)(4). CORRECTED INFORMATION: D 4. PRIMARY UDI NUMBER, H 3. DEVICE EVALUATED BY MANUFACTURER? ADDITIONAL INFORMATION: D 4. EXPIRATION DATE, H 4. DEVICE MANUFACTURE DATE, H 6. TYPE OF INVESTIGATION, H 6. INVESTIGATION FINDINGS, H 6. INVESTIGATION CONCLUSIONS, H 6. COMPONENT CODE. H3 INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO ETHICON FOR EVALUATION. VISUAL INSPECTION WAS CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT THE SECONDARY BOX OF THE VSTL45 PRODUCT CODE WITH LOT: 5881559 WAS RETURNED. UPON EVALUATION OF THE PACKAGING INSIDE OF THE SECONDARY BOX, THE PRIMARY PACKAGING OF PRODUCT CODE VSTL35 WITH THE RE-LABEL OF LOT: 5881559 WAS FOUND INSIDE. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE REPORTED COMPLAINT WAS CONFIRMED. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST MARKET SURVEILLANCE. A REVIEW OF THE DEVICE BATCH RECORD WAS PERFORMED FOR RAW MATERIALS, IN-PROCESS AND FINISHED GOODS FOR LOT NUMBER: 5881559. CAPA: 03781 WAS OPENED. THE DEVICES MET ALL RAW MATERIALS, IN-PROCESS AND FINISHED GOODS SPECIFICATIONS UPON RELEASE OF THE PRODUCT. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT # (B)(4). D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. H6. COMPONENT CODE: G07002 - DEVICE NOT RETURNED. ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ATTEMPTS ARE BEING MADE TO OBTAIN THE FOLLOWING INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. DID THIS ISSUE CONTRIBUTE TO ANY PATIENT ADVERSE EVENT? THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND BARBED SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 378856 | VISTASEAL LAPAROSCOPIC DUAL APPLICATOR (45 CM FLEXIBLE) | FIBRIN SEALANT PREPARATION DEVICE | MZM | ETHICON INC. | 5881559 | 10705031462724 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |