SERVO-S BASE UNIT
Report
- Report Number
- 8010042-2025-0001442
- Event Type
- Malfunction
- Date Received
- September 23, 2025
- Date of Event
- February 3, 2025
- Report Date
- September 23, 2025
- Manufacturer
- MAQUET CRITICAL CARE AB
- Product Code
- CBK
- PMA / PMN Number
- K123149
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
BASED ON PROVIDED INFORMATION, INSPECTION OF THE DEVICE WAS REQUESTED. THE PRESSURE TRANSDUCER BOARD, EXPIRATORY CHANNEL BOARD, AIR GAS MODULE, AC/DC CONVERTER AND EXPIRATORY CHANNEL CONNECTOR BOARD WERE IDENTIFIED AS DEFECTIVE. PRESSURE TRANSDUCER PRINTED CIRCUIT BOARD MEASURES THE PRESSURE, CONVEYED VIA THE PRESSURE TUBE CONNECTED TO THIS BLOCK BY ITS DIFFERENTIAL PRESSURE TRANSDUCER. WITH DIFFERENTIAL REFERENCE TO THE AMBIENT PRESSURE, THE OUTPUT SIGNAL IS PROPORTIONAL TO THE MEASURED PRESSURE THUS GIVING A LINEAR MEASUREMENT IN THE RANGE -40 CMH2O TO +160 CMH2O. EXPIRATORY CHANNEL BOARD CONTAINS ELECTRONICS INCLUDING MICROPROCESSOR FOR E.G. EXPIRATORY FLOW MEASUREMENT PERFORMED BY THE FLOWMETER INSIDE THE CASSETTE AND ELECTRONIC CONNECTIONS TO AND FROM THE CASSETTE, EXPIRATORY VALVE CONTROL, SAFETY VALVE CONTROL, TEMPERATURE MONITORING AND ALSO HOLDERS AND ELECTRICAL CONNECTORS FOR THE EXPIRATORY AND INSPIRATORY PRESSURE TRANSDUCER BOARDS. THE AC/DC CONVERTER IS A COMPLETE UNIT ADAPTED FOR THE SYSTEM. THE AC/DC CONVERTER CONVERTS THE CONNECTED AC POWER (MAINS POWER) TO THE INTERNAL DC SUPPLY VOLTAGE +12V_AC-DC. MAINS POWER IS SUPPLIED TO THE AC/DC CONVERTER VIA THE MAINS INLET. THE EXPIRATORY CHANNEL CONNECTOR IS A CONNECTOR BOARD, INCLUDING SIGNAL FILTERS. IT IS MOUNTED IN THE EXPIRATORY CASSETTE COMPARTMENT AND CONNECTS TO (B)(4) MOUNTED IN THE EXPIRATORY CASSETTE WHEN THE CASSETTE IS DOCKED TO THE EXPIRATORY CASSETTE COMPARTMENT. THE AIR/O2 GAS MODULE REGULATES THE INSPIRATORY GAS FLOW AND A FAULTY AIR/O2 GAS MODULE MAY AFFECT THE DELIVERED VOLUME OR GAS MIXTURE (O2/AIR). UNFORTUNATELY, HE DEVICE LOGS WERE NOT PROVIDED. THE CAUSE TO THE REPORTED ISSUES HAS NOT BEEN DETERMINED. THE UDI INFORMATION IS NOT AVAILABLE SINCE THE DEVICE WAS MANUFACTURED BEFORE 2015.
IT WAS REPORTED THAT THE PRESSURE TRANSDUCER BOARD, EXPIRATORY CHANNEL BOARD, AIR GAS MODULE, AC/DC CONVERTER AND EXPIRATORY CHANNEL CONNECTOR BOARD WERE DEFECTIVE. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: 1222205.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 532514 | SERVO-S BASE UNIT | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | MAQUET CRITICAL CARE AB | 6640440 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |