CONTACT DETACH
Report
- Report Number
- 3003442380-2025-14003
- Event Type
- Malfunction
- Date Received
- September 23, 2025
- Date of Event
- August 21, 2025
- Report Date
- October 3, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- PMA / PMN Number
- K041545
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS, H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6003437, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6003437 WAS MANUFACTURED ACCORDING TO THE VERSION 94 AND PACKAGING IN THE MULTIVAC M10 ON 28-SEP-2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING CONNECTOR OF THE LOT 4M03224 WAS MANUFACTURED ACCORDING TO THE WI VERSION 37 AND MANUFACTURED IN THE LINE L3, ON 18-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING CONNECTOR OF THE LOT 4M03226 WAS MANUFACTURED ACCORDING TO THE WI VERSION 37 AND MANUFACTURED IN THE LINE L3, ON 17-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING CONNECTOR OF THE LOT 4M03220 WAS MANUFACTURED ACCORDING TO THE[ WI VERSION 37 AND MANUFACTURED IN THE LINE L3, ON 22-DEC-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING CONNECTOR OF THE LOT 4M03222 WAS MANUFACTURED ACCORDING TO THE WI VERSION 37 AND MANUFACTURED IN THE LINE L3, ON 06-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING CONNECTOR OF THE LOT 4M03223 WAS MANUFACTURED ACCORDING TO THE WI VERSION 37 AND MANUFACTURED IN THE LINE L3, ON 07-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING CONNECTOR OF THE LOT 4M03226 WAS MANUFACTURED ACCORDING TO THE WI VERSION 37 AND MANUFACTURED IN THE LINE L3, ON 17-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING CONNECTOR OF THE LOT 4J04224 WAS MANUFACTURED ACCORDING TO THE WI VERSION 37 AND MANUFACTURED IN THE LINE L3, ON 22-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING CONNECTOR OF THE LOT 4J02557 WAS MANUFACTURED ACCORDING TO THE WI VERSION 37 AND MANUFACTURED IN THE LINE L3, ON 24-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2025. THE BLOCKAGE WAS AT THE SITE. THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 218470 | CONTACT DETACH | UNO CONTACT DETACH G29 60/6TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1002833 | 6003437 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR | Male |