FDA Adverse Event Malfunction Summary report: N

OPEN SPINE CLAMP

MDR report key: 23115655 · Received September 22, 2025

Report

Report Number
1723170-2025-03295
Event Type
Malfunction
Date Received
September 22, 2025
Date of Event
August 26, 2024
Report Date
September 22, 2025
Manufacturer
MEDTRONIC NAVIGATION, INC.
Product Code
OLO
PMA / PMN Number
K990214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3, H6: THE SPINE CLAMP, LOT NUMBER: 230227, WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE SCREW HOLE OF THE SPINE CLAMP WAS FOUND TO BE PARTIALLY STRIPPED THAT WOULD MAKE IT DIFFICULT TO TIGHTENED. THE REPORTED EVENT COULD BE DUPLICATED BY MEDTRONIC PERSONNEL. CODES B01, C07, AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT THE SCREW HOLE OF THE OPEN SPINE CLAMP WAS STRIPPED. THERE WAS NO REPORTED DELAY TO THE PROCEDURE DUE TO THIS ISSUE. THERE WAS NO REPORTED IMPACT ON PATIENT OUTCOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
218672 OPEN SPINE CLAMP ORTHOPEDIC STEREOTAXIC INSTRUMENT OLO MEDTRONIC NAVIGATION, INC. 9731780 230227

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown