FDA Adverse Event
Malfunction
Summary report: N
TRANSTAR STRETCHER
MDR report key: 2311180
·
Received October 20, 2011
Report
- Report Number
- 1824206-2011-05290
- Event Type
- Malfunction
- Date Received
- October 20, 2011
- Date of Event
- October 5, 2011
- Report Date
- October 5, 2011
- Manufacturer
- HILL-ROM INC
- Product Code
- FPO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN FOUND THE STRETCHER IN STORAGE TAGGED "BRAKE NOT WORKING ON HEAD END". FOUND SCREW BROKEN IN HEX ROD AND WELDMENT FOR B/S ROD BROKEN AT FRAME. REPLACED HEX ROD AND SCREW. DRILLED HOLES AND INSTALLED NEW BRAKE/STEER ROD BRACKET TO RESOLVE THIS ISSUE.
Description of Event or Problem · 1
STRETCHER FOUND IN STORAGE TAGGED "BRAKE NOT WORKING ON HEAD END". NO INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRANSTAR STRETCHER | WHEELED STRETCHER | FPO | HILL-ROM INC | P8000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |