FDA Adverse Event Malfunction Summary report: N

VERO 270 2.0ATA HARD SIDED SITTING HYPERBARIC CHAMBER

MDR report key: 23109677 · Received September 22, 2025

Report

Report Number
MW5176289
Event Type
Malfunction
Date Received
September 22, 2025
Date of Event
September 9, 2025
Report Date
September 16, 2025
Manufacturer
UNK
Product Code
CBF
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
NE, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED ON 01/21/2026 FOR REPORT MW5176289 TO UPDATE THE MANUFACTURER TO UNKNOWN IN THE FILE NAME. E3: UNK.

Description of Event or Problem · 0

VERO 270 2.0ATA HARD SIDED SITTING HYPERBARIC CHAMBER GOT CALLED ABOUT A HYPERBARIC CHAMBER INCIDENT OCCURRING ON (B)(6) 2025 WITH THE ABOVE MODEL. SOMEWHERE NEAR (B)(6). I TRACKED THE PATIENT DOWN AND SPOKE WITH HER. THE PATIENT HAD THE DOOR BLOW OFF HER CHAMBER THAT WAS IN HER HOME WHEN PRESSURIZED TO NEARLY AT NEARLY 2 ATA (14.7 PSIG). REPORTEDLY, THE ONLY INJURY WAS TO HER LEG FROM THE DOOR HITTING HER. SHE'S A NURSE PRACTITIONER TREATING HERSELF FOR MASTECTOMY WOUND RECOVERY FROM BREAST CANCER. SHE WANTS TO ADD THIS MODALITY TO HER OFFICE PRACTICE. THE DOOR BLEW OUT AT ~2ATA WITHOUT ISSUE EXCEPT BLUNT FORCE TO HER LEG FROM THE HATCH. SHE THINKS THE CHAMBER ATMOSPHERE WAS ABOUT 32% O2 AND 90% TO HER VIA MASK OR CANNULA.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED ON 01/12/2026 FOR REPORT MW5176289 TO UPDATE THE MANUFACTURER.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED ON 11/24/2025 FOR REPORT MW5176289 TO UPDATE THE MANUFACTURER. THIS IS TO REPLACE MY PRIOR REPORT (MW5176289 SUBMITTED ON 9/16/25), WHICH SOMEHOW LISTS THE WRONG MANUFACTURER AS TYPHOON WATER WARES. I WAS INFORMED BY THE PATIENT INVOLVED THAT SHE HAD PROCURED THE CHAMBER FROM A COMPANY IN OR NEAR (B)(6). THIS WAS PER A PHONE CONVERSATION WITH THE INVOLVED PATIENT. VERO 270 2.0ATA HARD SIDED SITTING HYPERBARIC CHAMBER GOT CALLED ABOUT A HYPERBARIC CHAMBER INCIDENT OCCURRING ON (B)(6) 2025 WITH THE ABOVE MODEL. SOMEWHERE NEAR (B)(6). I TRACKED THE PATIENT DOWN AND SPOKE WITH HER. THE PATIENT HAD THE DOOR BLOW OFF HER CHAMBER THAT WAS IN HER HOME WHEN PRESSURIZED TO NEARLY AT NEARLY 2 ATA (14.7 PSIG).REPORTEDLY, THE ONLY INJURY WAS TO HER LEG FROM THE DOOR HITTING HER. SHE'S A NURSE PRACTITIONER TREATING HERSELF FOR MASTECTOMY WOUND RECOVERY FROM BREAST CANCER. SHE WANTS TO ADD THIS MODALITY TO HER OFFICE PRACTICE. THE DOOR BLEW OUT AT ~2ATA WITHOUT ISSUE EXCEPT BLUNT FORCE TO HER LEG FROM THE HATCH. SHE THINKS THE CHAMBER ATMOSPHERE WAS ABOUT 32% O2 AND 90% TO HER VIA MASK OR CANNULA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
201671 VERO 270 2.0ATA HARD SIDED SITTING HYPERBARIC CHAMBER CHAMBER, HYPERBARIC CBF UNK VERO 270 2.0ATA HARD SIDED SITTI

Patients

Seq Age Sex Outcome Treatment
1 48 YR Female