FDA Adverse Event Malfunction Summary report: N

CARDELLA¿ PULMONARY ARTERY SENSOR SYSTEM

MDR report key: 23108776 · Received September 22, 2025

Report

Report Number
3024985933-2025-00030
Event Type
Malfunction
Date Received
September 22, 2025
Date of Event
August 29, 2025
Report Date
December 9, 2025
Manufacturer
ENDOTRONIX, INC
Product Code
MOM
PMA / PMN Number
P230040
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

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

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WILL BE SUBMITTED VIA FOLLOW-UP REPORT WITHIN 30 DAYS OF RECEIPT.

Description of Event or Problem · 0

THE PATIENT'S DATA WAS REVIEWED BY ENDOTRONIX'S INTERNAL MONITORING PROGRAM AND SUSPECTED SENSOR INACCURACY WAS IDENTIFIED. THE SITE HAS BEEN INFORMED TO SCHEDULE A RECALIBRATION. THE PATIENT IS CURRENTLY PENDING A RIGHT HEART CATHETERIZATION (RHC) AND SENSOR RECALIBRATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2535131 CARDELLA¿ PULMONARY ARTERY SENSOR SYSTEM PULMONARY ARTERY SENSOR MOM ENDOTRONIX, INC 100303-05 E230110-07

Patients

Seq Age Sex Outcome Treatment
1 77 YR Female