FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2310727 · Received October 24, 2011

Report

Report Number
1720753-2011-23170
Event Type
Malfunction
Date Received
October 24, 2011
Date of Event
October 12, 2011
Report Date
October 24, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND CLEANED AND RELUBRICATED THE HIGH VOLTAGE CABLE CONNECTORS. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM INTERMITTENTLY WOULD NOT DISPLAY AN IMAGE. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1