EASYPUMP
Report
- Report Number
- 9610825-2025-00549
- Event Type
- Malfunction
- Date Received
- September 20, 2025
- Date of Event
- August 4, 2025
- Report Date
- September 20, 2025
- Manufacturer
- B BRAUN MELSUNGEN AG
- Product Code
- MEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). REVIEWED THE DHR FOR BATCH NUMBERS 25A09GED81 AND 25B17GED81, THERE IS NO ABNORMALITY AND NO SUCH DEFECT DETECTED AT IN PROCESS AND AT FINAL CONTROL INSPECTION. NO SAMPLES RECEIVED AND NO MEANINGFUL PICTURE AVAILABLE FOR A PROPER EVALUATION. THE INVESTIGATION WAS ONLY DONE BASED ON COMPLAINT DESCRIPTION AND BATCH MANUFACTURING RECORD REVIEW AND HISTORICAL DATA. THERE ARE MULTIPLE POTENTIAL ROOT CAUSE FOR THIS DEFECT. 1. VALVE LEAKAGE: A CAPA HAS BEEN INITIATED TO ADDRESS THIS DEFECT. 2. PIN HOLE AT SILICON SLEEVE: CORRECTIVE ACTION ALREADY INITIATED AND COMPLETED IN ADDRESSING THIS DEFECT. 3. LEAK AT FILTER: A CAPA HAS BEEN INITIATED TO ADDRESS THIS DEFECT. AS NEITHER COMPLAINT SAMPLE NOR PICTURE WAS RECEIVED, FURTHER INVESTIGATION IS NOT POSSIBLE TO IDENTIFY THE LEAK. HENCE, THE COMPLAINT IS CLASSIFIED AS NOT CONFIRMED. THE COMPLAINT BATCH SHOWS NO ABNORMALITY. NOTE: THIS REPORTED IS BEING FILED FOR AN ITEM NUMBER THAT IS NOT SOLD IN THE UNITED STATES, HOWEVER SIMILAR ITEMS ARE SOLD IN THE UNITED STATES BY B. BRAUN MEDICAL, INC. PLEASE NOTE, THIS DEVICE IS DISTRIBUTED OUTSIDE THE US AND NO GUDID INFORMATION EXISTS. IT IS BEING REPORTED DUE TO A SIMILAR DEVICE BEING MARKETED WITHIN THE US WITH THE FOLLOWING INFO: UDI NUMBER (B)(4). PREMARKET SUBMISSION # K081905.
ACCORDING TO THE EVENT DESCRIPTION: "LEAK OF THE PUMP".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1991409 | EASYPUMP | PUMP, INFUSION, ELASTOMERIC | MEB | B BRAUN MELSUNGEN AG | 25A09GED81 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |