FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 23098149 · Received September 19, 2025

Report

Report Number
1645337-2025-10423
Event Type
Injury
Date Received
September 19, 2025
Date of Event
August 4, 2025
Manufacturer
MENTOR TEXAS
Product Code
FTR
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: SILENT RUPTURE MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. D4: UDI: AS THE SERIAL NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

SUSPECT DEVICE RECEIVED ON SEPTEMBER 22, 2025. DEVICE EVALUATION COMPLETED ON SEPTEMBER 28, 2025: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE BREAST IMPLANT HAD A TEAR EXTENDING FROM THE POSTERIOR VIEW TO THE ANTERIOR VIEW, MEASURING APPROXIMATELY 6.0 CM. IN ADDITION, SHELL ABRASION WAS OBSERVED AT THE EDGE OF THE RUPTURE ON THE POSTERIOR VIEW. THESE FINDINGS ARE CONSISTENT WITH NORMAL WEAR OF THE DEVICE. SHELL ABRASION SUGGESTS IN-VIVO FOLDING OR CREASING OF THE DEVICE, WHICH MAY BE CAUSED BY CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE, SUCH AS AN EXCESSIVELY SMALL BREAST POCKET OR THE FOLDING OR WRINKLING OF THE SHELL WITHIN THE BREAST POCKET. IN SOME CASES, BREAST IMPLANTS MAY ALSO EXPERIENCE NORMAL WEAR OVER TIME. THE CONTRALATERAL DEVICE WAS ALSO RECEIVED (LOT NUMBER 5623072). THE PATIENT DID NOT REPORT ANY ISSUE WITH THIS DEVICE. THEREFORE, NO FURTHER ANALYSIS OF THE CONTRALATERAL DEVICE IS REQUIRED. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA, AND THAT NORMAL WEAR MAY HAVE CAUSED THE OBSERVED DEFECT. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

A FEMALE PATIENT WHO UNDERWENT AN UNSPECIFIED BREAST AUGMENTATION WITH MENTOR MEMORYGEL, 450CC SILICONE PROSTHESIS EXPERIENCE RIGHT SIDED SILENT RUPTURE POST OPERATION. THE ISSUE WAS DIAGNOSED IN OFFICE AND CONFIRMED DURING SURGERY. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REPLACEMENTS AS FOLLOWS: LEFT SERIAL NUMBER 2003910-053, RIGHT SERIAL NUMBER (B)(6) -018 ON (B)(6) , 2025.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1475664 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 5567347

Patients

Seq Age Sex Outcome Treatment
1 66 YR Female Required Intervention