FDA Adverse Event
Malfunction
Summary report: N
QUANTUM INTERSEGMENTAL TABLE
MDR report key: 23095560
·
Received September 19, 2025
Report
- Report Number
- 3007278668-2025-00001
- Event Type
- Malfunction
- Date Received
- September 19, 2025
- Date of Event
- August 20, 2025
- Report Date
- September 16, 2025
- Manufacturer
- PIVOTAL HEALTH SOLUTIONS, INC.
- Product Code
- ISA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
THE DEVICE IS BEING RETURNED TO MANUFACTURER FOR EVALUATION. EVALUATION OF INITIAL INFORMATION DETERMINED IT WAS LIKELY THERE WAS MATERIAL BREAKDOWN OF HEAT COMPONENT CAUSED BY ACCELERATED WEAR AND REPEATED PINPOINT PRESSURE FROM PATIENT LOADING/UNLOADING RESULTING IN AN ELECTRICAL SHORT WITHIN THE INTERNAL COILS OF THE HEATING ELEMENT.
Description of Event or Problem · 0
A HEAT FUNCTION ON A QUANTUM TABLE MANUFACTURED IN 2010 MALFUNCTIONED AND OVERHEATED CAUSING THE VINYL TO MELT AND SMOLDER LEAVING A HOLE IN THE VINYL CUSHION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1397868 | QUANTUM INTERSEGMENTAL TABLE | QUANTUM | ISA | PIVOTAL HEALTH SOLUTIONS, INC. | Q400 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |