FDA Adverse Event Malfunction Summary report: N

QUANTUM INTERSEGMENTAL TABLE

MDR report key: 23095560 · Received September 19, 2025

Report

Report Number
3007278668-2025-00001
Event Type
Malfunction
Date Received
September 19, 2025
Date of Event
August 20, 2025
Report Date
September 16, 2025
Manufacturer
PIVOTAL HEALTH SOLUTIONS, INC.
Product Code
ISA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE IS BEING RETURNED TO MANUFACTURER FOR EVALUATION. EVALUATION OF INITIAL INFORMATION DETERMINED IT WAS LIKELY THERE WAS MATERIAL BREAKDOWN OF HEAT COMPONENT CAUSED BY ACCELERATED WEAR AND REPEATED PINPOINT PRESSURE FROM PATIENT LOADING/UNLOADING RESULTING IN AN ELECTRICAL SHORT WITHIN THE INTERNAL COILS OF THE HEATING ELEMENT.

Description of Event or Problem · 0

A HEAT FUNCTION ON A QUANTUM TABLE MANUFACTURED IN 2010 MALFUNCTIONED AND OVERHEATED CAUSING THE VINYL TO MELT AND SMOLDER LEAVING A HOLE IN THE VINYL CUSHION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1397868 QUANTUM INTERSEGMENTAL TABLE QUANTUM ISA PIVOTAL HEALTH SOLUTIONS, INC. Q400

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown