FDA Adverse Event Malfunction Summary report: N

INSET

MDR report key: 23093166 · Received September 19, 2025

Report

Report Number
3003442380-2025-13995
Event Type
Malfunction
Date Received
September 19, 2025
Date of Event
August 22, 2025
Report Date
September 30, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244016620
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6011714, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011714 WAS MANUFACTURED ACCORDING TO THE WI VERSION 120 AND MANUFACTURED IN THE LINE L6 ON 15-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. AN NC 2165520 FOR SEVERAL LOADS INTERRUPTED WAS PERFORMED FOR THE STERILIZATION PROCESS. THE SUB-ASSEMBLY: GLUING OF TUBING OF THE LOT 5B01679 WAS MANUFACTURED ACCORDING TO THE WI VERSION 66 AND MANUFACTURED IN THE MACHINE SC09, ON 11-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING OF TUBING OF THE LOT 5A06566 WAS MANUFACTURED ACCORDING TO THE WI VERSION 66 AND MANUFACTURED IN THE MACHINE SC04, ON 30-JAN-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: ONE NON-CONFORMANCE (NC) WAS RAISED DURING THE PROCESS UNRELATED TO THE MALFUNCTION REPORTED, THEREFORE, NO NC RAISED RELATED TO COMPLAINT CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED AN INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2025. THE BLOCKAGE WAS IN THE TUBING. BLOOD GLUCOSE LEVEL WAS DISPLAYED HIGH AT THE TIME OF THE EVENT. THEREFORE, PATIENT TOOK BOLUS TO RELOVE THE BLOOD GLUCOSE ISSUE. PATIENT CHANGED SUPPLIES AS NECESSARY AND RESUMED INSULIN THERAPY. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323886 INSET UNO INSET I 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1001680 6011714 05705244016620

Patients

Seq Age Sex Outcome Treatment
1 71 YR Male