FDA Adverse Event Malfunction Summary report: N

CONTACT DETACH

MDR report key: 23092752 · Received September 19, 2025

Report

Report Number
3003442380-2025-13949
Event Type
Malfunction
Date Received
September 19, 2025
Date of Event
August 21, 2025
Report Date
September 26, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244018457
PMA / PMN Number
K041545
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 01 - (B)(4). ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: PER REVISION 21 OF PROCEDURE (B)(4), A CHILD INVESTIGATION IS NOT REQUIRED TO DOCUMENT THE DEVICE HISTORY RECORD (DHR) REVIEW FOR A TYPE 2 REPORTABLE COMPLAINT. HOWEVER, THE CHILD WAS CREATED TO ALLOW THE PARENT RECORD TO ADVANCE TO REVIEW AND SUMMARY. SINCE IT WAS CREATED, THE DHR REVIEW WAS DOCUMENTED WITHIN THE CHILD. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6013089, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6013089 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 101 AND PACKAGING IN THE MACHINE MULTIVAC 14 ON 30-APR-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT AN EXTENDED INSPECTION WAS CREATED DUE TO 2D CODE FAILURE, EXTENDED INSPECTION WAS FOUND WITHIN SPECIFICATION CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR (B)(4) DEVICE 2 OF 3.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4) EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED THREE INFUSION SET EVENTS IN WHICH HE FORGOT TO REMOVE NEEDLE GUARD ON (B)(6) 2025. THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
269332 CONTACT DETACH UNO CONTACT DETACH G29 80/6TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1002834 6013089 05705244018457

Patients

Seq Age Sex Outcome Treatment
1 68 YR Female