SAFE-T-CENTESIS TRAY 6FR X 16CM
Report
- Report Number
- 1625685-2011-00026
- Event Type
- Malfunction
- Date Received
- October 26, 2011
- Date of Event
- September 23, 2011
- Report Date
- September 26, 2011
- Manufacturer
- CAREFUSION
- Product Code
- GCB
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). A COMPLAINT SAMPLE WAS NOT PROVIDED FOR EVALUATION. CONSEQUENTLY, THE QUALITY OF THE PRODUCT COULD NOT BE EVALUATED IN REGARDS TO THE REPORTED CONDITION. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED FOR THIS FAILURE MODE SINCE A COMPLAINT SAMPLE WAS NOT PROVIDED FOR EVALUATION. A REVIEW OF COMPLAINT DATA DID NOT IDENTIFY ANY TREND WITH THIS OR SIMILAR FAILURE MODES. A REVIEW OF APPLICABLE MANUFACTURING PROCEDURES AND QUALITY PLAN INSPECTIONS DID NOT IDENTIFY ANY ISSUES THAT MAY HAVE CONTRIBUTED TO THE REPORTED CONDITION. THE CATHETER/VALVE ASSEMBLY IS SUPPLIED BY AN EXTERNAL VENDOR (CAREFUSION - (B)(4)) AND IS NOT ALTERED BY THE (B)(4) FACILITY PRIOR TO PLACEMENT IN THE FINISHED GOOD TRAY. A DEVICE HISTORY REVIEW (DHR), RAW MATERIAL HISTORY FILES, AND THE STERILIZATION BATCH RECORDS FOR THE LISTED MANUFACTURING LOTS SHOWED NO RECORDED QUALITY PROBLEMS OR REJECTIONS RELATED TO THIS INCIDENT. THIS FAILURE MODE WILL BE ENTERED INTO THE SCAN TRACKING SYSTEM AND APPLICABLE NOTIFICATION WILL BE PROVIDED TO THE VENDOR OF THE CATHETER/VALVE ASSEMBLY (CAREFUSION - (B)(4)).
THE CATHETER WAS USED FOR A PARACENTESIS AND FLUID WAS LEAKING OUT FROM THE VALVE. THE PATIENT DID NOT HAVE AN INJURY. SAMPLE NOT AVAILABLE. ON (B)(6) 2011, THE RADIOLOGIST INVOLVED INDICATED THAT DURING A PARACENTESIS, THE VALVE ON THE TOP OF THE CATHETER WAS LEAKING SO WAS ASPIRATING AIR WHEN CONNECTED TO DRAIN THE FLUID. THE RADIOLOGIST CONFIRMED THERE WAS NO INJURY TO THE PATIENT OR MEDICAL INTERVENTION REQUIRED. ON (B)(6) 2011, THE LAB MANAGER ((B)(6)) INDICATED THAT THE RADIOLOGIST HAD NOT PUNCTURED THE VALVE MORE THAN ONCE AS HE IS AN EXPERIENCED, SKILLED, USER OF THIS PRODUCT. THE LAB MANAGER ALSO INDICATED THAT AN UNUSED SAMPLE FROM THE SAME LOT IS NOT AVAILABLE. NO FURTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SAFE-T-CENTESIS TRAY 6FR X 16CM | NEEDLE, CATHETER | GCB | CAREFUSION | PIG1260T | 0000341522 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |