FDA Adverse Event Malfunction Summary report: N

SOFTBANK II

MDR report key: 2308754 · Received September 27, 2011

Report

Report Number
1058332-2011-00004
Event Type
Malfunction
Date Received
September 27, 2011
Report Date
September 16, 2011
Manufacturer
SCC SOFT COMPUTER
Product Code
MMH
Removal / Correction Number
1058332-09/16/2011-011-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RE: THE CLIENT REPORTED THE PRODUCT MALFUNCTION ON SOFTBANK II VERSION 25.1.0.2. AFFECTED PRODUCTS INCLUDE SOFTBANK II VERSIONS 23.2.0.1-23.2.0.5, 25.0.0.3-25.0.0.5, 25.1.0.1-25.1.0.2.

Description of Event or Problem · 1

WHEN PRINTING AN ISBT LABEL FOR A BLOOD PRODUCT WITH AN EXPIRATION TIME OF MIDNIGHT (00:00), THE GENERATED LABEL PRINTS 23:59 INSTEAD AS THE EXPIRATION TIME IN THE BARCODE. THE UNIT CAN BE MISLABELED WITH A FALSELY EXTENDED EXPIRATION TIME. NO ADVERSE PT EVENT IS ASSOCIATED WITH THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SOFTBANK II MMH SCC SOFT COMPUTER

Patients

Seq Age Sex Outcome Treatment
1