POWERED WHEELCHAIR
Report
- Report Number
- 1525712-2011-00635
- Event Type
- Malfunction
- Date Received
- October 26, 2011
- Date of Event
- September 29, 2011
- Report Date
- October 25, 2011
- Manufacturer
- INVACARE TAYLOR STREET
- Product Code
- ITI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
RMA HAS NOT BEEN INITIATED FOR THIS ISSUE. MODEL TDSP-CG (B)(4) IS APPROXIMATELY 3 YEARS OF AGE. THE USER MANUAL PART NUMBER 1143190 WAS ISSUED WITH THIS DEVICE. IT IS UNKNOWN IF THE CONSUMER HAS FULLY READ AND UNDERSTANDS THE USER MANUAL. DOCUMENTATION PROVIDES WARNINGS, CAUTIONS, AND INSTRUCTIONS FOR SAFELY USING THE DEVICE. IF THE CONSUMER DOES NOT UNDERSTAND THE WRITTEN WARNINGS, CAUTIONS OR INSTRUCTIONS THEN THEY SHOULD CONTACT INVACARE. THE CONSUMERS MEDICAL CONDITION, STABILITY AND MEDICATION REGIMEN ARE UNKNOWN . THE CONSUMERS AGE, HEIGHT AND WEIGHT ARE UNKNOWN. THE CONSUMERS TECHNIQUE WHILE USING THE DEVICE IS UNKNOWN. THE MAINTENANCE HISTORY OF THE DEVICE IS UNKNOWN.
(B)(4) ISSUED MFG REPORT #1525712-2011-00635. PRODUCT WAS RETURNED FOR AN EVALUATION. MOTOR WAS VERIFIED AS LEAKING GREASE. THE RISK ASSOCIATED WITH FAILURES OF THIS TYPE WAS EVALUATED UNDER (B)(4). THE RISK WAS DETERMINED TO BE AT AN ACCEPTABLE LEVEL. THIS INCIDENT DOES NOT IMPACT THAT ASSESSMENT. THIS MALFUNCTION IS CONSIDERED A RELIABILITY ISSUE THAT IS NOT LIKELY TO CAUSE HARM TO THE USER AND HAS BEEN ADDRESSED BY ENGINEERING. QUALITY REPORTS FOR CORRECTIVE ACTIONS HAVE BEEN PUT IN PLACE AND ARE EFFECTIVE IN ELIMINATING THE LEAKING GREASE CONDITION. RMA HAS NOT BEEN INITIATED FOR THIS ISSUE. MODEL TDSP-CG SERIAL NUMBER/DATE CODE (B)(4) IS APPROXIMATELY 3 YEARS OF AGE. THE USER MANUAL PART NUMBER 1143190 WAS ISSUED WITH THIS DEVICE. IT IS UNKNOWN IF THE CONSUMER HAS FULLY READ AND UNDERSTANDS THE USER MANUAL. DOCUMENTATION PROVIDES WARNINGS, CAUTIONS, AND INSTRUCTIONS FOR SAFELY USING THE DEVICE. IF THE CONSUMER DOES NOT UNDERSTAND THE WRITTEN WARNINGS, CAUTIONS OR INSTRUCTIONS THEN THEY SHOULD CONTACT INVACARE. THE CONSUMER'S MEDICAL CONDITION, STABILITY AND MEDICATION REGIMEN ARE UNKNOWN . THE CONSUMER'S AGE, HEIGHT AND WEIGHT ARE UNKNOWN. THE CONSUMER'S TECHNIQUE WHILE USING THE DEVICE IS UNKNOWN. THE MAINTENANCE HISTORY OF THE DEVICE IS UNKNOWN.
BACK CANES ARE ALLEGEDLY BENT. LEFT MOTOR GEARBOX IS ALLEGEDLY MAKING A NOISE. NO INJURY IS ALLEGED.
BACK CANES ARE ALLEGEDLY BENT. LEFT MOTOR GEARBOX IS ALLEGEDLY MAKING A NOISE. NO INJURY IS ALLEGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | POWERED WHEELCHAIR | 890.3860 | ITI | INVACARE TAYLOR STREET | TDXSP-CG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |