POWERPICC
Report
- Report Number
- 3006260740-2025-07018
- Event Type
- Malfunction
- Date Received
- September 17, 2025
- Date of Event
- September 3, 2025
- Report Date
- November 7, 2025
- Manufacturer
- C.R. BARD, INC. (BASD) -3006260740
- Product Code
- LJS
- UDI-DI
- 00801741138966
- PMA / PMN Number
- UNK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. THE COMPLAINT OF A DAMAGED CATHETER WAS CONFIRMED. THE PRODUCT RETURNED FOR EVALUATION WAS ONE 4 FR SINGLE LUMEN POWERPICC CATHETER. THE INVESTIGATION FINDINGS ARE CONSISTENT WITH DAMAGE ACCUMULATED THROUGH FLEXURAL FATIGUE. FLEXURAL FATIGUE MAY OCCUR DUE TO CYCLIC KINKING OF THE CATHETER TUBE IN WHICH PHYSIOLOGICAL, PLACEMENT, USAGE, AND MECHANICAL FACTORS MAY GRADUALLY FORM A CRACK(S) IN THE CATHETER. THE RETURNED PRODUCT SAMPLE WAS EVALUATED, AND A SMALL SPLIT WAS OBSERVED BETWEEN THE 5 CM AND 6 CM DEPTH MARKERS AND ALSO A SECOND SPLIT WAS OBSERVED BETWEEN THE 8 CM AND 9 CM DEPTH MARKERS. THE CATHETER SPLITS CONTAINED PHYSICAL FEATURES ASSOCIATED WITH MATERIAL FATIGUE, AND THE CHARACTERISTICS OBSERVED WHICH SUPPORTED THIS TYPE OF FAILURE INCLUDED: ¿ DAMAGE WHICH WAS CIRCUMFERENTIALLY ALIGNED. ¿ FRACTURE EDGES WHICH WERE ROUNDED AND POLISHED DUE TO REPEATED MATERIAL WEAR. ¿ 'C' SHAPED IMPRESSIONS LEADING INTO THE FRACTURE SITE WHICH ARE CONSISTENT WITH MATERIAL BUCKLING DUE TO MOVEMENT WHICH CAUSED THE FRACTURE EDGES TO BE PRESSED TOGETHER. ¿ OVERALL ELLIPTICAL SHAPE TO THE FRACTURE CROSS-SECTION (A RESULT OF REPEATED KINKING OF THE TUBING). REPETITIVE KINKING OF THE INDWELLING CATHETER APPEARS TO HAVE CONTRIBUTED TO THE OBSERVED LEAK; HOWEVER, THE SPECIFIC CIRCUMSTANCES SURROUNDING HOW KINKING DEVELOPED INTO A FRACTURE WITHIN THE CATHETER TUBING WERE ULTIMATELY UNKNOWN. THIS COMPLAINT WILL BE RECORDED FOR FUTURE TRENDING AND MONITORING PURPOSES.
H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. EACH EVENT REPORTED TO BD IS EVALUATED AND INVESTIGATED IN ACCORDANCE WITH OUR COMPLAINT INVESTIGATION PROCEDURES. THE INVESTIGATION PROCESS INCLUDES, BUT IS NOT LIMITED TO, EVALUATION OF THE EVENT DETAILS PROVIDED BY THE COMPLAINT FACILITY, A REVIEW OF COMPLAINT HISTORY, MANUFACTURING RECORDS AND RISK DOCUMENT WHERE APPLICABLE, AND AN EVALUATION OF THE SUBJECT DEVICE WHEN AVAILABLE TO IDENTIFY POTENTIAL CONTRIBUTING FACTORS.
IT WAS REPORTED THAT A PICC WAS REMOVED WITH A PORE AT 8 CM. PATIENT HAS UNDERGONE 4 CYCLES OF QMT TREATMENT (WEEKLY CISPLATIN). ARM CIRCUMFERENCE: 38 CM VEIN CALIBER: 4.1 MM DEPTH: 2 CM BASILICA VEIN CANNULATED. 38 CM OF CATHETER INSERTED. SECURACATH USED FOR FIXATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2668314 | POWERPICC | PERIPHERALLY INSERTED CENTRAL CATHETER (PICC) | LJS | C.R. BARD, INC. (BASD) -3006260740 | N/A | REJX1835 | 00801741138966 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |