ACCESS® 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-04848
- Event Type
- Malfunction
- Date Received
- October 25, 2011
- Date of Event
- October 4, 2011
- Report Date
- October 4, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K922823/A007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
BECKMAN COULTER INC. ASSESSMENT OF CUSTOMER SUPPLIED INSTRUMENT PERFORMANCE DATA INDICATED THAT LEVEL TWO FT4 QUALITY CONTROL (QC) RESULTS WERE BELOW THE CUSTOMER'S ESTABLISHED RANGES BY TWO STANDARD DEVIATIONS ON THE MORNING OF THE EVENT. UPON REPEAT OF QC AFTER ASSAY RECALIBRATION, QC RESULTS RECOVERED WITHIN RANGE. A DEFINITIVE ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED TO DATE.
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUS FREE THYROXINE (FT4) PATIENT RESULTS IN CONJUNCTION WITH FREE T4 QUALITY CONTROL PERFORMANCE OUTSIDE OF THEIR ESTABLISHED RANGES ON AN ACCESS 2 IMMUNOASSAY SYSTEM. BECKMAN COULTER INC. ASSESSMENT OF CUSTOMER SUPPLIED DATA REVEALED THE GENERATION OF NINE INITIALLY ERRONEOUS PATIENT RESULTS. THE INITIAL PATIENT RESULTS WERE ASSOCIATED WITH A TIMEFRAME WHEN TWO FREE T4 QUALITY CONTROL ASSESSMENTS FAILED TO PERFORM WITHIN THE CUSTOMER'S ESTABLISHED LIMITS. OTHER ASSAY PATIENT RESULTS WERE ALSO INCLUDED IN SUPPLIED DATA, HOWEVER THE CUSTOMER WAS NOT QUESTIONING THESE RESULTS. AFTER RE-CALIBRATION OF THE ASSAY, ACCEPTABLE FT4 QC RESULTS WERE GENERATED. PATIENT REPEAT TESTING WAS COMMENCED AND YIELDED RESULTS THAT THE CUSTOMER DID NOT QUESTION AS ERRONEOUS. BECKMAN COULTER INC. ASSESSMENT OF CUSTOMER SUPPLIED DATA INDICATED THAT FOUR OF THE INITIAL FT4 RESULTS WERE BELOW THE ASSAY'S NORMAL REFERENCE RANGE, WITH THE REMAINDER OF RESULTS FALLING WITHIN THE ASSAY'S NORMAL REFERENCE RANGE. UPON REPEAT TESTING ON THE SAME INSTRUMENT, SIX PATIENT SAMPLES CROSS CLINICAL RANGES. ALL REPEAT RESULTS, WITH EXCEPTION OF TWO, RECOVERED WITHIN THE ASSAY'S NORMAL REFERENCE RANGE. THESE TWO REPEAT RESULTS RECOVERED ABOVE THE NORMAL REFERENCE RANGE. THREE PATIENT INITIAL AND REPEAT SAMPLE RESULTS COLLECTIVELY EXCEEDED THE LABELED FREE T4 ASSAY PRECISION CLAIMS. THE ERRONEOUS FT4 RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY AND HENCE THERE HAVE BEEN NO REPORTS OF DEATH, SERIOUS INJURY OR MODIFICATION TO PATIENT TREATMENT ASSOCIATED OR ATTRIBUTED TO THIS EVENT. PATIENT SPECIFIC INFORMATION AND SAMPLE HANDLING/COLLECTION INFORMATION WERE NOT PROVIDED BY THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS® 2 IMMUNOASSAY SYSTEM | ANALYZER, CHEMISTRY | JJE | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ACCESS FREE T4 REAGENT |