FDA Adverse Event Death Summary report: N

HEATER-COOLER SYSTEM

MDR report key: 23072809 · Received September 17, 2025

Report

Report Number
9611109-2025-90361
Event Type
Death
Date Received
September 17, 2025
Date of Event
August 7, 2025
Report Date
September 17, 2025
Manufacturer
LIVANOVA DEUTSCHLAND GMBH
Product Code
DWC
UDI-DI
04033817901099
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

A.1.-A.5. THERE WAS NO PATIENT INVOLVEMENT. G.5. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K220635). H11: LIVANOVA DEUTSCHLAND MANUFACTURES THE A LIVANOVA FIELD SERVICE ENGINEER WAS DISPATCHED THE CUSTOMER AND INSPECTED THE DEVICE. THE DISTRIBUTION BOARD AND PATIENT PUMP 2 WILL LIKELY BE REPLACED TO SOLVE THE ISSUE. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 0

LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT THE 3T HEATER COOLER DISPLAYED AN ALARM MEANING PUMP 1 IS BLOCKED (TOO SLOW, E07) DURING DISINFECTION. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2553201 HEATER-COOLER SYSTEM CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND GMBH HEATER COOLER 3T 04033817901099

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown