FDA Adverse Event Malfunction Summary report: N

MATRIXMANDIBLE 1.5MM DRILL BIT J-LATCH W/8MM STOP/50MM

MDR report key: 23066891 · Received September 16, 2025

Report

Report Number
8030965-2025-09398
Event Type
Malfunction
Date Received
September 16, 2025
Date of Event
August 27, 2025
Manufacturer
SYNTHES GMBH
Product Code
HWE
UDI-DI
10887587012889
PMA / PMN Number
K043310
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL NARRATIVE: H3, H6: PART: 03.503.408 SYNTHES LOT: U400221 SUPPLIER LOT: U400221 RELEASE TO WAREHOUSE DATE: 24 JUNE 2022 SUPPLIER: (B)(4). NO NONCONFORMANCE REPORTS (NCRS) WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE PRODUCT WAS NOT RETURNED TO MEDTECH ORTHOPAEDICS, HOWEVER PHOTOS WERE RECEIVED FOR REVIEW. THE PHOTO INVESTIGATION REVEALED THAT '03.503.408, MATRIXMAND 1.5 DB J-LATCH W/8MM STOP/50¿ HAD BROKEN. THE OBSERVED CONDITION OF THE DEVICE WAS CONSISTENT WITH A RANDOM COMPONENT FAILURE THAT MAY HAVE BEEN CAUSED BY EXPOSURE TO UNINTENDED FORCES. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE MATRIXMAND 1.5 DB J-LATCH W/8MM STOP/50 WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, POTENTIAL CAUSE CAN BE ATTRIBUTED TO COMPONENT FAILURE AND IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

DURING THE PROCEDURE, WHILE THE SPECIALIST WAS PERFORMING THE DRILLING, THE DRILL BIT BROKE. THE DRILL BIT ALONG WITH ITS FRAGMENTS WERE REMOVED. ANOTHER DRILL BIT WITH THE SAME CHARACTERISTICS WAS PASSED TO THE DOCTOR, THUS SUCCESSFULLY COMPLETING THE SURGERY. THERE WAS NO HARM TO THE PATIENT AND THERE WERE NO ALTERATIONS IN SURGICAL TIMES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
276226 MATRIXMANDIBLE 1.5MM DRILL BIT J-LATCH W/8MM STOP/50MM INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT HWE SYNTHES GMBH U400221 10887587012889

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown