UNICEL DXC 800 SYNCHRON SYSTEM
Report
- Report Number
- 2050012-2011-06534
- Event Type
- Malfunction
- Date Received
- October 24, 2011
- Date of Event
- September 25, 2011
- Report Date
- September 25, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE ROOT CAUSE OF THE INSTRUMENT ISSUE COULD NOT BE DETERMINED; HOWEVER, THE REPAIRS AND ADJUSTMENTS EFFECTIVELY RESOLVED THE ISSUE. (NOTE: THE BECKMAN COULTER, INC. IDENTIFIER FOR THIS REPORT IS (B)(4).)
CUSTOMER CALLED TO REPORT THAT THE UNICEL DXC 800 SYNCHRON SYSTEM GENERATED FALSELY HIGH MAGNESIUM (MG) RESULTS AND ERRATIC GLUCOSE (GLU) RESULTS. CUSTOMER REPORTED THAT A TECHNICIAN RAN THREE MAGNESIUM PATIENTS SAMPLES, ALL OF WHICH YIELDED RESULTS OF 6.0 MILLIGRAMS PER DECILITER (MG/DL). THE TECHNICIAN REPEATED THE PATIENT SAMPLES ON ANOTHER INSTRUMENT AND RECEIVED RESULTS OF 2.0 MG/DL, 2.0 MG/DL, AND 1.9 MG/DL. ALSO, THE INSTRUMENT GENERATED A CRITICALLY LOW GLUCOSE RESULT OF 39 MG/DL, FOLLOWED BY A REPEATED RESULT OF 53 MG/DL; THE OTHER INSTRUMENT GENERATED A RESULT OF 40 GM/DL. ON THE FOLLOWING DAY, THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT AND FOUND THAT THE SAMPLE MIXER WASH STATION WAS OBSTRUCTED, CAUSING THE SAMPLE MIXER TO STAY UNWASHED. THE REAGENT MIXER WASH STATION WAS PARTIALLY OBSTRUCTED, PRODUCING ONE WASHING STREAM. THE FSE REMOVED AND THEN CLEANED THE MIXER WASH STATION INSERTS. THE FSE THEN VERIFIED THAT THE MIXER WASH STATIONS WERE FUNCTIONING PROPERLY. THE FSE ALSO RE-ALIGNED THE SAMPLE MIXER PADDLE, THE WASH COLLAR, AND THE PROBE ROTARY. LASTLY, THE FSE REPLACED THE SAMPLE PROBE. CUSTOMER DID NOT PROVIDE SPECIFIC PATIENT INFORMATION. CUSTOMER STATED THAT NO ERRONEOUS PATIENT RESULTS WERE REPORTED OUTSIDE THE LABORATORY; THEREFORE, PATIENT TREATMENT WAS NOT AFFECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 800 SYNCHRON SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER, INC. | DXC 800 PRO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |