RONIN X6 OR HEADLIGHT

MDR report key: 23054567 · Received September 15, 2025

Report

Report Number
MW5176076
Event Type
Malfunction
Date Received
September 15, 2025
Date of Event
September 3, 2025
Report Date
September 10, 2025
Manufacturer
RONIN SURGICAL CORP.
Product Code
HPP
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
NJ, US
Reporter Occupation
RISK MANAGER
Health Professional
Yes

Narratives

Description of Event or Problem · 0

DURING A SURGICAL PROCEDURE IN THE OPERATING ROOM, THE RONIN X6 HEADLIGHT WAS USED. TWO SURGEONS WORE HEADLIGHTS THROUGHOUT THE APPROXIMATELY TWO-HOUR SURGERY. UPON COMPLETION OF THE PROCEDURE, IT WAS OBSERVED THAT THE PATIENT'S ABDOMEN--SPECIFICALLY THE AREA EXPOSED OUTSIDE THE STERILE DRAPE--APPEARED DISCOLORED AND SHOWED SIGNS OF SMALL BLISTERING. A CONSULTATION WITH THE PLASTIC SURGERY TEAM WAS CONDUCTED, AND THEY CONFIRMED THAT THE AFFECTED AREA SUSTAINED A BURN INJURY. PATIENT WAS A 13-MONTH-OLD CHILD. BIO ENGINEERING TESTED EVERY RONIN HEADLIGHT THAT WE HAVE ONSITE. NOTHING WAS FOUND TO BE WRONG WITH THE HEADLIGHTS. PEAK INTENSITY: 350,000 LUX ADJUSTABLE SPOT DIAMETER: 5.6" TO 1.5" AT 18" WORKING DISTANCE RUNTIME: 5+ HOURS (EXTENDED BATTERY), 2.5 HOURS (STANDARD BATTERY) WEIGHT: 6.2 OZ LUMINAIRE WIRELESS DESIGN WITH HAPTIC, AUDIBLE, AND VISUAL BATTERY INDICATORS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2640318 RONIN X6 OR HEADLIGHT HEADLAMP, OPERATING, BATTERY-OPERATED HPP RONIN SURGICAL CORP. RK0350

Patients

Seq Age Sex Outcome Treatment
1 13 MO Female Other