FDA Adverse Event Malfunction Summary report: N

ARCHITECT TOTAL B-HCG

MDR report key: 23037562 · Received September 12, 2025

Report

Report Number
3005094123-2025-00450
Event Type
Malfunction
Date Received
September 12, 2025
Date of Event
September 1, 2025
Report Date
December 1, 2025
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
DHA
UDI-DI
00380740163297
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW, AND IN HOUSE ACCURACY TESTING. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE WITHOUT INDICATION FOR ANY ADDITIONAL ISSUE. A REVIEW OF TRACKING AND TRENDING DID IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY. HOWEVER, NO COMMONALITIES FOR THE COMPLAINT LOT AND ISSUE WERE IDENTIFIED. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES, POTENTIAL NON-CONFORMANCES OR DEVIATIONS ASSOCIATED WITH THE COMPLAINT LOT OR COMPLAINT ISSUE. ACCURACY TESTING WAS COMPLETED WITH A RETAINED KIT OF THE COMPLAINT LOT. ALL VALIDITY AND ACCEPTANCE CRITERIA WERE MET, AND THE PRODUCT IS PERFORMING AS EXPECTED. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER¿S ISSUE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ARCHITECT TOTAL B-HCG REAGENT, LOT 71474UD02 WAS IDENTIFIED.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION E1 - INITIAL REPORTER ESTABLISHMENT NAME COMPLETE ENTRY = (B)(6). ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 07K78, THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US WITH THE SAME LIST NUMBER, AND A 510K/PMA/BLA NUMBER OF K983424.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSE POSITIVE ARCHITECT TOTAL -HCG RESULTS GENERATED FOR A 23-YEAR-OLD FEMALE PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED (REFERENCE RANGE <5 MIU/ML IS NEGATIVE): SAMPLE ID (B)(6), INITIAL RESULT = 38.23 MIU/ML, REPEAT RESULT = 37.76 MIU/ML COLLOIDAL GOLD RESULT = NEGATIVE ROCHE RESULT = <0.2 MIU/ML PEG TREATMENT RESULT = 40.14 MIU/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSE POSITIVE ARCHITECT TOTAL -HCG RESULTS GENERATED FOR A 23-YEAR-OLD FEMALE PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED (REFERENCE RANGE <5 MIU/ML IS NEGATIVE): SAMPLE ID (B)(6), INITIAL RESULT = 38.23 MIU/ML, REPEAT RESULT = 37.76 MIU/ML, COLLOIDAL GOLD RESULT = NEGATIVE, ROCHE RESULT = <0.2 MIU/ML, PEG TREATMENT RESULT = 40.14 MIU/ML . NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
127871 ARCHITECT TOTAL B-HCG SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN DHA ABBOTT IRELAND DIAGNOSTICS DIVISION 71474UD02 00380740163297

Patients

Seq Age Sex Outcome Treatment
1 23 YR Female ARC I2000SR INST, 03M74-02, (B)(6)| ARC I2000SR INST, 03M74-02, (B)(6)