FDA Adverse Event Injury Summary report: N

QUICK SET

MDR report key: 23035332 · Received September 12, 2025

Report

Report Number
3003442380-2025-13622
Event Type
Injury
Date Received
September 12, 2025
Date of Event
August 8, 2025
Report Date
September 18, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244017450
PMA / PMN Number
K991759
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010396, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 17-SEP-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL "6010396". THE COUNT OF COMPLAINT IS 1 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010396 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 82 AND MANUFACTURED IN THE MACHINE 12 ON 21-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. NON-CONFORMANCE (NC) LOW TEMPERATURE IN HEATING AEREATION PHASE AT (B)(4) DURING STERILIZATION PROCESS. NO RELATION WITH MALFUNCTION CODE. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: NO PHOTO WAS PROVIDED. IN ORDER TO TEST THE PRODUCT, THE RETURNED SAMPLE FROM THE LOT HAVE BEEN REQUESTED. NO RETURNED SAMPLES. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED. ONE NC RAISED DURING PRODUCTION NO RELATED TO COMPLAINT CODE, ONE COMPLAINT THRESHOLDS IDENTIFIED FOR THE LOT IN QUESTION AND SERIOUS INJURY/DEATH, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Additional Manufacturer Narrative · 0

E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: BRAZIL.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN BRAZIL. IT WAS REPORTED THAT THE PATIENT FACED DIABETIC KETOACIDOSIS EVENT ON (B)(6) 2025 AND EVENTUALLY GOT HOSPITALIZED. SINCE INFUSION SET TUBING GOT BENT. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2330529 QUICK SET UNO QUICK-SET 60/6 SC1 MECA FPA UNOMEDICAL DEVICES S.A. DE C.V. MMT-399A 6010396 05705244017450

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown