ROCHE 9180 ELECTROLYTE ANALYZER
Report
- Report Number
- 1823260-2025-02804
- Event Type
- Malfunction
- Date Received
- September 9, 2025
- Date of Event
- August 20, 2025
- Report Date
- November 17, 2025
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JFP
- UDI-DI
- 04015630031832
- PMA / PMN Number
- K961458
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
FOR THE PATIENT WITH A MONTH LONG ABSENCE OF BOWEL MOVEMENTS, ALSO KNOWN AS SEVERE CONSTIPATION, DOESN'T DIRECTLY AFFECT BLOOD CHLORIDE (CL) OR SODIUM (NA+) LEVELS. ELECTROLYTE IMBALANCES ARE NOT A TYPICAL FEATURE OF CONSTIPATION. FOR THE PATIENT WITH DIARRHEA, THE CONDITION IS MORE LIKELY TO CAUSE ELECTROLYTE IMBALANCES. THE INVESTIGATION CONFIRMED THAT THE REPORTED ISSUE WITH ERRONEOUS SODIUM (NA) RESULTS WAS CAUSED BY A DEFECTIVE NA ELECTRODE (LOT# 31250347). THE ANALYSIS OF THE RETURNED PART SHOWED NO PHYSICAL DEFECTS BUT CRITICALLY REVEALED THAT THE INNER ELECTROLYTE SOLUTION WAS COMPLETELY DRIED OUT. THIS WOULD PREVENT THE ELECTRODE FROM FUNCTIONING CORRECTLY. THE K ELECTRODE (LOT# 31245047) AND REFERENCE ELECTRODE (LOT# 200843) FROM THE SAME RETURN WERE FOUND TO BE FREE OF DEFECTS. INTERMITTENT FLUIDIC OBSTRUCTIONS ARE THE MOST LIKELY ROOT CAUSE FOR THE OBSERVED ISSUE WITH THE CHLORIDE ELECTRODE. A REVIEW OF ANALYZER ALARM DATA SHOWED MULTIPLE OBSTRUCTION ERRORS. AN OBSTRUCTION IN THE FLUIDIC PATH (NEEDLE, INLET PORT) WOULD CAUSE INCONSISTENT SAMPLE ASPIRATION, LEADING TO THE NON-REPRODUCIBLE RESULTS SEEN IN BOTH PATIENT SAMPLES AND QC. THIS ALIGNS WITH THE FIELD SERVICE ENGINEER'S ACTION OF DECONTAMINATING THE NEEDLE AND INLET PORT. A REVIEW OF THE CERTIFICATES OF ANALYSIS WAS PERFORMED FOR THE PROVIDED ELECTRODE LOT NUMBERS. THE DOCUMENTATION CONFIRMS THAT ALL COMPONENTS PASSED THE INTERNAL MANUFACTURING AND QUALITY CONTROL CHECKS, MEETING ALL REQUIRED SPECIFICATIONS. NO PRODUCT ISSUE IS FOUND. THE RECOMMENDED WORKAROUND IS TO REPLACE THE FAULTY ELECTRODE AND MAKE SURE THE INNER ELECTROLYTE IS SUFFICIENT.
THE SODIUM ELECTRODE LOT NUMBER IS 31250347, WITH AN INSTALL BEFORE DATE OF 12-SEP-2025. THE CHLORIDE ELECTRODE LOT NUMBER WAS 31250547, WITH AN INSTALL BEFORE DATE OF 15-AUG-2025. THE FIELD SERVICE ENGINEER INSPECTED THE INSTRUMENT. THE ELECTRODES WERE REPLACED AND THE REFERENCE HOUSING WAS DEPROTEINIZED. THE NEEDLE AND INLET PORT WERE DECONTAMINATED. DAILY MAINTENANCE WAS PERFORMED. CALIBRATION AND CONTROLS RAN CORRECTLY. A PATIENT SAMPLE WAS PROCESSED SIX TIMES AND THE RESULTS WERE REPRODUCIBLE. A CALIBRATION REPORT PROVIDED FROM 22-AUG-2025 WAS ACCEPTABLE. THE INVESTIGATION IS ONGOING.
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT SODIUM ELECTRODE RESULTS FOR TWO PATIENT SAMPLES TESTED ON A ROCHE 9180 ELECTROLYTE ANALYZER. THE SECOND SAMPLE ALSO HAD DISCREPANT CHLORIDE ELECTRODE RESULTS. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE FIRST PATIENT SAMPLE RESULTED IN SODIUM VALUES OF 140 MMOL/L, 192.6 MMOL/L, AND 146.3 MMOL/L WITH A DATA FLAG. THE SECOND PATIENT SAMPLE RESULTED IN SODIUM VALUES OF 137.8 MMOL/L, 161.4 MMOL/L WITH A DATA FLAG, AND 173.1 MMOL/L WITH A DATA FLAG. THIS SAMPLE RESULTED IN CHLORIDE VALUES OF 108.8 MMOL/L, 96.2 MMOL/L WITH A DATA FLAG, AND 97.2 MMOL/L.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2827646 | ROCHE 9180 ELECTROLYTE ANALYZER | ELECTROLYTE ANALYZER | JFP | ROCHE DIAGNOSTICS | 04015630031832 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |