PLATINIUM
Report
- Report Number
- 1000165971-2025-00590
- Event Type
- Malfunction
- Date Received
- September 9, 2025
- Date of Event
- July 1, 2025
- Report Date
- December 1, 2025
- Manufacturer
- MICROPORT CRM S.R.L.
- Product Code
- NIK
- UDI-DI
- 08031527014524
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ANALYSIS OF THE AVAILABLE PATIENT FILES REVEALED: O AN ABNORMAL BATTERY DEPLETION O NO OVERCONSUMPTION - THE EXPERTISE OF THE RETURNED CRTD DIDN¿T REVEAL OVER-CONSUMPTION. THE BATTERY ANALYSIS REVEALED 2 CLUSTERS WERE FOUND INSIDE THE BATTERY. BOTH CLUSTERS WERE IN CONTACT WITH THE CATHODE BRIDGE, AND LEAD TO THE FAST BATTERY DEPLETION. - THE BATTERY FAILURE IS THE ROOT CAUSE OF THE FAST BATTERY DEPLETION.
ANALYSIS OF THE AVAILABLE PATIENT FILES REVEALED: O AN ABNORMAL BATTERY DEPLETION O NO OVERCONSUMPTION - THE EXPERTISE OF THE RETURNED CRTD DIDN'T REVEAL OVER-CONSUMPTION. - THE ROOT CAUSE COULD NOT BE DETERMINED WITH CERTAINTY. THE MOST LIKELY HYPOTHESIS WOULD BE A BATTERY FAILURE. FURTHER INVESTIGATIONS ARE ONGOING AT THE BATTERY MANUFACTURER LEVEL.
REPORTEDLY, A PREMATURE BATTERY DEPLETION IS SUSPECTED. THE SUBJECT DEFIBRILLATOR IS NOT INTERROGABLE ANYMORE. ON THE REMOTE MONITORING REPORT DATED ON (B)(6) 2025, THE BATTERY VOLTAGE WAS 2.79V AND 2.03 EARLY AUGUST.
REPORTEDLY, A PREMATURE BATTERY DEPLETION IS SUSPECTED. THE SUBJECT DEFIBRILLATOR IS NOT INTERROGABLE ANYMORE. ON THE REMOTE MONITORING REPORT DATED 8TH JULY 2025, THE BATTERY VOLTAGE WAS 2.79V AND 2.03 EARLY (B)(6).
REPORTEDLY, A PREMATURE BATTERY DEPLETION IS SUSPECTED. THE SUBJECT DEFIBRILLATOR IS NOT INTERROGABLE ANYMORE. ON THE REMOTE MONITORING REPORT DATED 8TH JULY 2025, THE BATTERY VOLTAGE WAS 2.79V AND 2.03 EARLY AUGUST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2068053 | PLATINIUM | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MICROPORT CRM S.R.L. | PLATINIUM 4LV SONR CRT-D 1844 | 08031527014524 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Hospitalization| R |