O-ARM O2 IMAGING SYSTEM
Report
- Report Number
- 3004785967-2025-00603
- Event Type
- Malfunction
- Date Received
- September 9, 2025
- Date of Event
- August 14, 2025
- Report Date
- October 3, 2025
- Manufacturer
- MEDTRONIC NAVIGATION, INC. (BOXBOROUGH)
- Product Code
- OWB
- UDI-DI
- 00643169639683
- PMA / PMN Number
- K200074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
(H3,H6): THE SOFTWARE INVESTIGATION FOUND THAT THE REPORTED EVENT WAS NOT A SOFTWARE ISSUE. COMPLAINT DATA ANALYSIS FOUND THE EVENT IS DUE TO A USER WORKFLOW ISSUE AS PER THE COMPLAINT DATA. VERIFIED ACCURATE IMAGING SYSTEM NAVIGATION AND PERFORMED IMAGING SYSTEM CHECKOUT. COULD NOT RECREATE THE NAVIGATION INACCURACY. UNIT WAS PROPERLY FUNCTIONING AND HAD BEEN RETURNED TO SERVICE. SOFTWARE FUNCTIONING AS DESIGNED. B01,C19,D14 CODES APPLICABLE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
H2) ADDITIONAL INFORMATION WAS ADDED TO E. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
(H3, H6): MANUFACTURING REPRESENTATIVE WENT TO THE SITE TO TEST THE SYSTEM, VERIFIED ACCURATE IMAGING SYSTEM NAVIGATION AND PERFORMED IMAGING SYSTEM CHECKOUT. COULD NOT RECREATE THE NAVIGATION INACCURACY. UNIT WAS PROPERLY FUNCTIONING AND HAS BEEN RETURNED TO SERVICE. CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: (B)(6), SERIAL/LOT #:(B)(6), MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IMF CODE ADDED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
MEDTRONIC RECEIVED INFORMATION REGARDING AN IMAGING SYSTEM BEING USED FOR A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT DURING A 3D SPIN, THE IMAGES COME ACROSS AS INACCURATE. THE IMAGES WERE VIEWED ON THE MOBILE VIEWING STATION (MVS) AND HAD NOT YET BEEN SENT TO THE NAVIGATION SYSTEM AND THE IMAGES FAIL TO UPLOAD DICOM. NO FURTHER INFORMATION WAS PROVIDED REGARDING THE INACCURACY. THIS HAD OCCURRED INTRA OPERATIVELY. THERE WAS NO REPORTED IMPACT TO PATIENT OUTCOME. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1915042 | O-ARM O2 IMAGING SYSTEM | INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM | OWB | MEDTRONIC NAVIGATION, INC. (BOXBOROUGH) | BI70002000 | - | 00643169639683 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | "SEE H11." |