FDA Adverse Event Injury Summary report: N

IMP,TSV,MCOL MG,4.7MM,10M

MDR report key: 23001367 · Received September 9, 2025

Report

Report Number
0002023141-2025-02453
Event Type
Injury
Date Received
September 9, 2025
Report Date
October 17, 2025
Manufacturer
ZIMMER DENTAL
Product Code
DZE
PMA / PMN Number
K111889
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). A1: PATIENT IDENTIFIER (B)(6). A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED. D4: UNIQUE IDENTIFIER (UDI) NUMBER NOT AVAILABLE. E1: EMAIL ADDRESS UNKNOWN / NOT PROVIDED. G4: PREMARKET IDENTIFICATION K111889 AND K101880.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: TYPE OF INVESTIGATION CODE WAS ADDED: 10, 4109, 3331 AND 4111. H6: INVESTIGATION FINDINGS CODE WAS ADDED: 3252. H6: INVESTIGATION CONCLUSIONS CODE WAS ADDED: 4307. H10: NARRATIVE/DATA WAS UPDATED. ZIMVIE RECEIVED THE REPORTED DEVICE FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICE IDENTIFIED THE IMPLANT WITH SIGNS OF USE, APPARENT BONE / TISSUE ATTACHED TO EXTERNAL THREADS. THE IMPLANT WAS OBSERVED FRACTURED AT THE COLLAR REGION. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE AND EVENT. THIS COMPLAINT REFERS TO THE SPECIFIC DEVICE BEING INVESTIGATED FOR THIS COMPLAINT RECORD. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. THE CUSTOMER DID NOT SUBMIT IMAGES FOR THE REPORTED EVENT. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE LONG-TERM PARAFUNCTIONAL HABITS (E.G., CLENCHING, BRUXISM, AND OVERLOADING) OF THE PATIENT OVER THE IMPLANTATION PERIOD ¿ PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE REPORTED EVENT WAS CONFIRMED WITH ALL THE AVAILABLE INFORMATION. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE FRACTURED PROSTHETIC REVERED BEVEL. ABUTMENT GOT LOOSE AFTER FRACTURED REVERSED BEVEL OF THE PROSTHETIC TABLE. IMPLANT COLLAR FRACTURED, NO ISSUE WITH ABUTMENT PER DISCUSSION WITH DOCTORS ASSISTANT. IMPLANT WAS REMOVED. ADDITIONAL APPOINTMENT REQUIRED TO REPLACE A NEW IMPLANT. BONE TYPE: 2 . TOOTH SITE: 31.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2066956 IMP,TSV,MCOL MG,4.7MM,10M DENTAL IMPLANT DZE ZIMMER DENTAL 62546412

Patients

Seq Age Sex Outcome Treatment
1 84 YR Male Required Intervention