CONTACT DETACH
Report
- Report Number
- 3003442380-2025-13512
- Event Type
- Malfunction
- Date Received
- September 9, 2025
- Date of Event
- August 11, 2025
- Report Date
- October 13, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244018426
- PMA / PMN Number
- K041545
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED EXPIRATION DATE UNDER D4 AND MANUFACTURING DATE UNDER H4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: REVISION 21 OF (B)(4) DOES NOT REQUIRE A COMPLAINT THAT IS TYPE 2 REPORTABLE TO OPEN A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS OPENED AGAINST A PREVIOUS REVISION OF (B)(4). COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4) THE BATCH 6012846, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6012846 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 101 AND PACKAGING IN THE MACHINE MULTIVAC 14 ON 13-APR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, WELDING OF THE LOT 5D01362 WAS MANUFACTURED ACCORDING TO THE WI VERSION 34 AND MANUFACTURED IN THE MACHINE LS24-LS25, ON 12-APR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING CONNECTOR OF THE LOT 5D01336 WAS MANUFACTURED ACCORDING TO THE WI VERSION 39 AND MANUFACTURED IN THE MACHINE GLUING 3, ON 12-APR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING CONNECTOR OF THE LOT 5D02703 WAS MANUFACTURED ACCORDING TO THE WI VERSION 39 AND MANUFACTURED IN THE MACHINE GLUING 3, ON 12-APR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING CONNECTOR OF THE LOT 5D02704 WAS MANUFACTURED ACCORDING TO THE WI VERSION 39 AND MANUFACTURED IN THE MACHINE GLUING 3, ON 12-APR-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
REFERENCE NUMBER: (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKED ALARM EVENT ON 11-AUG-2025. THE BLOCKAGE WAS AT THE SITE. THE BLOOD GLUCOSE LEVELS WAS HIGH AND THE PATIENT TREATED WITH CORRECTION INJECTION VIA MULTIPLE DAILY INJECTION (MDI). THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2067879 | CONTACT DETACH | UNO CONTACT DETACH G29 60/6TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1002833 | 6012846 | 05705244018426 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 YR | Female |