PRISMAFLEX SETS (ST)
Report
- Report Number
- 8010182-2025-00542
- Event Type
- Malfunction
- Date Received
- September 8, 2025
- Date of Event
- August 5, 2025
- Report Date
- October 6, 2025
- Manufacturer
- VANTIVE US HEALTHCARE LLC
- Product Code
- KDI
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
E1: INITIAL REPORTER FACILITY NAME:(B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
H11: THE ACTUAL DEVICE WAS NOT AVAILABLE; HOWEVER, PHOTOGRAPHS OF THE SAMPLE WERE PROVIDED FOR EVALUATION. VISUAL INSPECTION OF THE PHOTOS DID NOT SHOW THE REPORTED LEAK FROM THE SET ACCESS POST-PUMP POD BUT SHOWED THAT THE "PRIME SELF-TEST (CODE 2)" ALARM WAS TRIGGERED ON THE MACHINE. THE COMPONENT AFFECTED BY THE LEAK, THE ACCESS POST-PUMP POD, IS MANUFACTURED BY A VANTIVE EXTERNAL SUPPLIER. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE POD DEFECT WAS DETERMINED TO BE RELATED TO SUPPLIER MANUFACTURING, AND THE ISSUE IS BEING FURTHER INVESTIGATED. NONCONFORMANCE RECORDS HAVE BEEN OPENED TO ADDRESS THIS ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT AN EXTERNAL FLUID LEAK WAS OBSERVED FROM THE FILTER PRESSURE SENSOR OF A PRISMAFLEX ST100 SET DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1915756 | PRISMAFLEX SETS (ST) | DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM | KDI | VANTIVE US HEALTHCARE LLC | NA | 24E0061 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | N/A. |