VALKYRIE THORACIC FIXATION SYSTEM
Report
- Report Number
- 3014680795-2025-00033
- Event Type
- Injury
- Date Received
- September 8, 2025
- Date of Event
- August 19, 2025
- Report Date
- September 4, 2025
- Manufacturer
- J.M. LONGYEAR MANUFACTURING, LLC
- Product Code
- HRS
- PMA / PMN Number
- K232515
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- 003
Narratives
THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR PART 803. THIS REPORT MAY BE BASED ON INFORMATION THAT HAS NOT BEEN ABLE TO BE COMPLETELY INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION THAT THE REPORT CONSTITUTES AND ADMISSION THAT THE DEVICE, THE COMPANY, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. THE PRODUCTS HAVE NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION, THEREFORE A DHR REVIEW COULD NOT BE CONDUCTED. A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED. IF ADDITIONAL INFORMATION RELEVANT TO THE INVESTIGATION OR OTHER CONTENT OF THIS REPORT IS IDENTIFIED, THIS REPORT WILL BE UPDATED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 3014680795-2025-00034; 3014680795-2025-00035; 3014680795-2025-00036.
IT WAS REPORTED A PATIENT THAT UNDERWENT A STERNAL CLOSURE USING A VARIETY OF PLATES AND SCREWS ALONG WITH FIBER TAPE ON (B)(6) 2025 EXPERIENCED A STERNAL DEHISCENCE THAT CORRELATED WITH EXTREME COUGHING FITS. THE PATIENT UNDERWENT REVISION SURGERY TO CORRECT THE STERNAL DEHISCENCE. DURING REOPERATION IT WAS IDENTIFIED THAT ONE X-PLATE DISSOCIATED FROM THE BONE AND THE SECOND X-PLATE COMPLETELY FRACTURED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2066408 | VALKYRIE THORACIC FIXATION SYSTEM | X-PLATE | HRS | J.M. LONGYEAR MANUFACTURING, LLC | 01-020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Male | Hospitalization| R |