FDA Adverse Event Malfunction Summary report: N

CD HORIZON SOLERA 5.5/6.0

MDR report key: 22992030 · Received September 8, 2025

Report

Report Number
3003120897-2025-00189
Event Type
Malfunction
Date Received
September 8, 2025
Date of Event
August 13, 2025
Report Date
November 4, 2025
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC.
Product Code
HXX
UDI-DI
00643169043640
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS OF PART# 5584111 LOT# K23D1185 VISUAL AND OPTICAL EXAMINATION IDENTIFIED IT APPEARS THAT PART OF THE TIP OF THE INSTRUMENT HAS BEEN BROKEN OFF. THE METAL DEFORMATION GIVES INDICATION THAT THE FAILURE WAS THE RESULT OF OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING PRODUCTS USED FOR SPINAL THERAPY. IT WAS REPORTED THAT MULTIPLE INSTRUMENTS WERE FOUND TO BE BROKEN OR WORN DUE TO NORMAL WEAR AND TEAR. A BROKEN TIP, RATCHETING HANDLE WAS WORN DOWN AND DOES NOT RATCHET, ANGLE POSITIONER DEVICE WAS WORN AND WOULD NOT ENGAGE THE SCREW HEAD, AND AN OBTURATOR THAT WAS STRIPPED AT THE END. THERE WAS NO PATIENT INVOLVED AND NO FURTHER COMPLICATIONS OR SYMPTOMS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2056442 CD HORIZON SOLERA 5.5/6.0 SCREWDRIVER HXX MEDTRONIC SOFAMOR DANEK USA, INC. 5584111 K23D1185 00643169043640

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown