TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM
Report
- Report Number
- 8030965-2025-09047
- Event Type
- Malfunction
- Date Received
- September 5, 2025
- Date of Event
- August 14, 2025
- Manufacturer
- SYNTHES GMBH
- Product Code
- JEY
- UDI-DI
- 07611819908376
- PMA / PMN Number
- K083388
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H6: PHOTO INVESTIGATION: THE PRODUCT WAS NOT RETURNED TO J&J MEDTECH ORTHOPAEDICS; HOWEVER, PHOTOS WERE PROVIDED FOR REVIEW. THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED TO J&J MEDTECH ORTHOPEDICS FOR EVALUATION. THE PHOTOGRAPH ATTACHED WAS REVIEWED, THE HEAD SCREW CAN BE OBSERVED. HOWEVER, THE IMAGE QUALITY IS POOR AND NO OBSERVATIONS PERTAINING TO THE NATURE OF THE REPORTED EVENT COULD BE IDENTIFIED. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE PHOTOGRAPHS ATTACHED ARE INSUFFICIENT TO DRAW A CONCLUSION ABOUT THE REPORTED EVENT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. H3, H4, H6: DEVICE HISTORY RECORD (DHR) REVIEW CONDUCTED: PART# 04.503.206.01C, LOT # 41845P7, MANUFACTURING SITE: DEPUY SYNTHES PRODUCTS, INC, SUPPLIER: (B)(4), RELEASE TO WAREHOUSE DATE: 08 NOV 2024, EXPIRATION DATE: NA. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW WAS BROKEN DURING INSERTION. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THERE WAS NO SURGICAL DELAY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1976981 | TI MATRIXMIDFACE SCREW SELF-TAPPING 6MM | PLATE, BONE | JEY | SYNTHES GMBH | 41845P7 | 07611819908376 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |