FDA Adverse Event Injury Summary report: N

DXU 840M IRIS AUTOMATED URINE MICROSCOPY ANALYZER

MDR report key: 22977423 · Received September 5, 2025

Report

Report Number
2122870-2025-00095
Event Type
Injury
Date Received
September 5, 2025
Date of Event
August 9, 2025
Report Date
September 5, 2025
Manufacturer
BECKMAN COULTER IRELAND, INC.
Product Code
KQO
UDI-DI
15099590748654
PMA / PMN Number
K022774
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE BECKMAN FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE NOTED THAT THE RED CHECK VALVE IN THE FRONT LEFT OF THE INSTRUMENT WAS BROKEN; ONE OF THE TABS HOLDING THE TWO HALVES OF THE VALVE TOGETHER HAD BROKEN OFF. TO RESOLVE THE ISSUE, THE FSE REPLACED THE RED CHECK VALVE. THE OPERATOR WAS ASSESSED BY THEIR OCCUPATIONAL HEALTH DEPARTMENT THE OPERATOR HAD BRUISED BOTH KNEES, BRUISED AN ELBOW, AND REPORTED SHOULDER PAIN. THE OPERATOR WAS TAKING PAIN MEDICATION; HOWEVER, THE CUSTOMER DID NOT DIVULGE WHETHER THE PAIN MEDICATION WAS PRESCRIBED OR OVER THE COUNTER. ADDITIONALLY, THE OPERATOR WAS ICING THE AFFECTED AREAS AND RESTING. WITH THE INFORMATION AVAILABLE IT COULD NOT BE CONFIRMED WHETHER OR NOT MEDICAL INTERVENTION IN THE FORM OF PRESCRIBED MEDICATION WAS REQUIRED. THIS REPORT HAS BEEN SUBMITTED IN AN ABUNDANCE OF CAUTION. BECKMAN COULTER INTERNAL IDENTIFIER (B)(4).

Description of Event or Problem · 0

THE CUSTOMER¿S DXU 840M IRIS INSTRUMENT HAD AN UNCONTAINED LEAK RESULTING IN AN OPERATOR SLIPPING AND FALLING TO THE FLOOR. THE OPERATOR WAS WEARING THE PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) HOWEVER THEY CAME INTO CONTACT WITH THE LEAK UPON FALLING. THE LEAK CONSISTED OF LAMINA. THERE WAS NO DIRECT CONTACT TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE OPERATOR WHO HAD SLIPPEDL ON THE LIQUID ON THE FLOOR WAS EXPECTED TO REPORT TO THEIR SITE'S OCCUPATIONAL HEALTH. NO ERRONEOUS PATIENT RESULTS WERE RELEASED OUT FROM THE CUSTOMER¿S LABORATORY IN RELATION TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1975656 DXU 840M IRIS AUTOMATED URINE MICROSCOPY ANALYZER DXU 840M IRIS KQO BECKMAN COULTER IRELAND, INC. INSTRUMENT DXU 840M IRIS 15099590748654

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other