FDA Adverse Event Injury Summary report: N

TECNIS ODYSSEY TORIC II IOL

MDR report key: 22968455 · Received September 4, 2025

Report

Report Number
3012236936-2025-000231
Event Type
Injury
Date Received
September 4, 2025
Date of Event
April 12, 2025
Report Date
January 9, 2026
Manufacturer
AMO MANUFACTURING NETHERLANDS
Product Code
MFK
UDI-DI
05050474811409
PMA / PMN Number
P980040
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SECTION H3: THE DEVICE WAS NOT RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. A REVIEW OF RECORDS INCLUDING DEVICE HISTORY AND COMPLAINT TRENDING WILL BE PERFORMED. UPON COMPLETION OF THE REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

UPON FURTHER REVIEW IT WAS NOTED THAT SECTION G4: PMA/510(K) NUMBER IN THE INITIAL MDR WAS INADVERTENTLY SUBMITTED MISSING THE ¿P¿ BEFORE THE NUMERICAL VALUE. THE COMPLETE NUMBER IS P980040. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: SECTION D9: DEVICE AVAILABLE FOR EVALUATION: YES. SECTION D9: DEVICE DATE RETURNED TO MANUFACTURER: SEPT 29, 2025. SECTION H3: DEVICE EVALUATED BY MANUFACTURER: YES. DEVICE EVALUATION: VISUAL INSPECTION OF THE COMPLAINT DEVICE REVEAL THAT THE LENS WAS CUT IN HALF AND COATED IN VISCOELASTIC RESIDUE. THE LENS WAS CLEANED AND INSPECTED AND NO FURTHER ISSUES WERE IDENTIFIED. THE COMPLAINT ISSUES WERE NOT IDENTIFIED DURING PRODUCT EVALUATION. BASED ON THE PRODUCT EVALUATION, THE COMPLAINT ISSUES COULD NOT BE CONFIRMED TO BE RELATED TO A MANUFACTURING OR DESIGN ISSUE. THEREFORE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY OR PRODUCT MALFUNCTION. BASED ON THE COMPLAINT INVESTIGATION RESULTS, THE PRODUCT WAS RELEASED WITHIN SPECIFICATIONS. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED BY THE CUSTOMER THAT A DIB00 +20.0 INTRAOCULAR LENS (IOL) WAS EXPLANTED DUE TO DYSPHOTOPSIA AND BLURRED VISION IN THE LEFT EYE, CAUSING DIFFICULTY WITH DRIVING AND READING. THE LENS WAS REMOVED AND REPLACED DURING A SECONDARY SURGICAL PROCEDURE WITH AN IOL FROM ANOTHER COMPANY. NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2082361 TECNIS ODYSSEY TORIC II IOL LENS, MULTIFOCAL INTRAOCULAR MFK AMO MANUFACTURING NETHERLANDS DRT150 05050474811409

Patients

Seq Age Sex Outcome Treatment
1 60 YR Female Required Intervention