FDA Adverse Event Malfunction Summary report: N

BD PHOENIX¿ PMIC/ID-107

MDR report key: 22963621 · Received September 4, 2025

Report

Report Number
1119779-2025-04918
Event Type
Malfunction
Date Received
September 4, 2025
Date of Event
October 1, 2024
Report Date
July 10, 2025
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
LON
UDI-DI
30382904486076
PMA / PMN Number
K020322
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT IS UNKNOWN. B3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. D.4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H.4. DEVICE MANUFACTURE DATE: UNKNOWN. G4. MULTIPLE 510(K)#: K021954, K023273, K023301, K024152, K030677, K031306, K031679, K031679, K032131, K033784, K033907, K040006, K040106, K040716, K050089, K050555, K051689, K053241, K060214, K060217, K060218, K060493, K070809, K082538, K082852, AND K131331. INVESTIGATION SUMMARY: THIS COMPLAINT IS FOR MISIDENTIFICATION WHEN USING PHOENIX PANEL PMIC/ID-107 (CATALOG NUMBER 448607) BATCH NUMBER UNKNOWN. THE CUSTOMER DID NOT RETURN PANELS, ISOLATES OR PHOENIX GENERATED LAB REPORTS FOR THE INVESTIGATION. THE BATCH NUMBER WAS NOT PROVIDED; THEREFORE, THIS COMPLAINT IS UNCONFIRMED. A REVIEW OF QUALITY NOTIFICATIONS COULD NOT BE PERFORMED SINCE A BATCH NUMBER WAS NOT PROVIDED. COMPLAINT TRENDING WAS PERFORMED, AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. PLEASE CONTINUE TO COMMUNICATE ANY ADDITIONAL CONCERNS. THIS MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING (B)(6) 2023-2025, UNDER CAPA 11910483

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD PHOENIX¿ PMIC/ID-107, A PATIENT SAMPLE WAS INCORRECTLY IDENTIFIED. THERE WAS NO REPORT OF PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2254833 BD PHOENIX¿ PMIC/ID-107 SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION LON BECTON DICKINSON & CO. (SPARKS) UNKNOWN 30382904486076

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown