FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 22954442 · Received September 3, 2025

Report

Report Number
9617229-2025-14932
Event Type
Injury
Date Received
September 3, 2025
Date of Event
June 3, 2025
Report Date
September 3, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
UDI-DI
10888628033412
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF CAPSULAR CONTRACTURE WAS RECEIVED ON OCTOBER 07, 2025, WITH LOT NUMBER 1252303. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS ARE: ¿ CAPSULAR CONTRACTURE: UNABLE TO OBSERVE AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. AS PER THE INVESTIGATION PROCEDURES, DEFORMATION AND CREASES WERE COMPLETED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE GRADE 3.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE HAS BEEN EXPLANTED AND REPLACED. TREATMENT: SINGULAIR 10 MG.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE HAS BEEN EXPLANTED AND REPLACED. TREATMENT: SINGULAIR 10 MG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1386622 INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 1252303 10888628033412

Patients

Seq Age Sex Outcome Treatment
1 42 YR Female Required Intervention