FDA Adverse Event Malfunction Summary report: N

SELECTSECURE MRI SURESCAN

MDR report key: 22953135 · Received September 3, 2025

Report

Report Number
2182208-2025-04391
Event Type
Malfunction
Date Received
September 3, 2025
Date of Event
May 28, 2025
Report Date
September 3, 2025
Manufacturer
MEDTRONIC, INC.
Product Code
NVN
PMA / PMN Number
P030036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS INFORMATION IS BASED ENTIRELY ON JOURNAL LITERATURE. MEDTRONIC WAS MADE AWARE OF THIS EVENT THROUGH A SEARCH OF LITERATURE PUBLICATIONS. THIS EVENT OCCURRED OUTSIDE THE US. PATIENT INFORMATION IS LIMITED DUE TO CONFIDENTIALITY CONCERNS. OF NOTE, MULTIPLE PATIENTS AND MULTIPLE MANUFACTURERS WERE NOTED IN THE ARTICLE; HOWEVER, A ONE-TO-ONE CORRELATION COULD NOT BE MADE WITH UNIQUE PRODUCT SERIAL/LOT NUMBERS. THE BASELINE GENDER/AGE CHARACTERISTICS IS MALE/11 YEARS OLD. WITHOUT A LOT NUMBER OR DEVICE SERIAL NUMBER, THE MANUFACTURING DATE CANNOT BE DETERMINED. SINCE NO DEVICE ID WAS PROVIDED, IT IS UNKNOWN IF THIS EVENT HAS BEEN PREVIOUSLY REPORTED. A REQUEST FOR ADDITIONAL INFORMATION WAS MADE. IF ADDITIONAL INFORMATION IS OBTAINED REGARDING THIS EVENT, IT WILL BE ADDED, AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY. REFERENCED ARTICLE: TRANSVENOUS PACING LEADS IMPLANTED IN EPICARDIAL POSITION: A RECIPE FOR FUTURE EPICARDIAL PACING ELECTRODE? JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY (JTCVS) TECHNIQUES. 2025. 32:144-51. DOI: 10.1016/J.XJTC.2025.05.011 MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

LITERATURE WAS REVIEWED REGARDING TRANSVENOUS PACING LEADS IMPLANTED IN EPICARDIAL POSITION. THE AUTHORS DESCRIBED ONE RIGHT VENTRICULAR (RV) LEAD WHICH EXHIBITED AN INCREASE IN THRESHOLD AND UNSTABLE SENSING MEASUREMENTS AND WAS EXPLANTED AND REPLACED. THERE WAS ALSO ONE RIGHT ATRIAL (RA) LEAD WITH AN INCREASE IN PACING THRESHOLDS. THE STATUS OF THE LEADS IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS OR PRODUCT PERFORMANCE ISSUES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2256111 SELECTSECURE MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) NVN MEDTRONIC, INC. 383069

Patients

Seq Age Sex Outcome Treatment
1 11 YR Male UNKNOWN COMPETITOR IPG