FDA Adverse Event Malfunction Summary report: N

BD VERITOR¿ SYSTEM FOR RAPID DETECTION OF SARS-COV-2

MDR report key: 22951935 · Received September 3, 2025

Report

Report Number
3006948883-2025-00646
Event Type
Malfunction
Date Received
September 3, 2025
Date of Event
July 4, 2024
Report Date
July 10, 2025
Manufacturer
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
Product Code
QKP
UDI-DI
00382902560821
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

G.4. PMA/510(K)#: EUA# (B)(4). INVESTIGATION SUMMARY THIS STATEMENT SUMMARIZES THE INVESTIGATION RESULTS REGARDING A COMPLAINT THAT ALLEGES FALSE NEGATIVE WHEN USING BD VERITOR¿ SARS-COV-2 KIT (MATERIAL#: 256082), BATCH NUMBER 4081455. THE CUSTOMER REPORTED THAT THEY SAW THE CONTROL LINE AND THE TEST LINE ON THE CARTRIDGE BUT RECEIVED A NEGATIVE RESULT FROM ANALYZER WHEN THEY TESTED A PATIENT WITH SYMPTOMS. BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE ALL COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, VISUAL INSPECTION OF RETENTION SAMPLES, AND VISUAL INSPECTION OF CUSTOMER RETURNED SAMPLES, IF APPLICABLE. BATCH HISTORY RECORD (BHR) REVIEW AND RETAIN SAMPLE TESTING WERE PERFORMED ON THE BATCH NUMBER PROVIDED. THE RESULTS WERE ACCEPTABLE, AND NO RELEVANT ISSUES WERE FOUND. NO PHOTOS WERE RECEIVED. RETURN SAMPLES WERE RECEIVED ON THE BATCH NUMBER PROVIDED. THEY WERE TESTED AND THE RESULTS WERE ACCEPTABLE. THE REPORTED ISSUE WAS UNABLE TO BE CONFIRMED. A TREND ANALYSIS FOR FALSE NEGATIVE WAS CONDUCTED; NO ADVERSE TREND WAS IDENTIFIED. THERE WERE NO CORRECTIVE ACTIONS TAKEN AT THIS TIME. THIS MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING (B)(6) 2023-2025, UNDER CAPA 11910483.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VERITOR¿ SYSTEM FOR RAPID DETECTION OF SARS-COV-2, THERE WAS A FALSE NEGATIVE RESULT. THERE WAS NO REPORT OF PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2256037 BD VERITOR¿ SYSTEM FOR RAPID DETECTION OF SARS-COV-2 CORONAVIRUS ANTIGEN DETECTION SYSTEM QKP BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. 4081455 00382902560821

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown