FDA Adverse Event Malfunction Summary report: N

BD PHOENIX¿ PID

MDR report key: 22933450 · Received September 2, 2025

Report

Report Number
1119779-2025-04744
Event Type
Malfunction
Date Received
September 2, 2025
Date of Event
August 21, 2024
Report Date
July 10, 2025
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
LQL
UDI-DI
30382904480081
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY THIS COMPLAINT IS FOR MISIDENTIFICATION OF STAPHYLOCOCCI WHEN USING PHOENIX PANEL PID (CATALOG NUMBER 448008) BATCH NUMBERS 4072152 AND 4121406. THE CUSTOMER DID NOT RETURN PANELS OR ISOLATES BUT PROVIDED PHOENIX GENERATED LAB REPORTS FOR THE INVESTIGATION. THE CUSTOMER PROVIDED PHOENIX LAB REPORTS SHOW AN ISOLATE IDENTIFIED AS STAPHYLOCOCCUS AUREUS, STAPHYLOCOCCUS EPIDERMIDIS, AND STAPHYLOCOCCUS LUGDENENSIS WHEN USING COMPLAINT BATCHES 4072152 AND 4121406. TO INVESTIGATE, RETENTION PANELS FROM THE COMPLAINT BATCHES WERE TESTED USING IN HOUSE AND QC ISOLATES S. AUREUS A43300 AND S. SAPROPHYTICUS POS 432 ON A PHOENIX M50 MACHINE AND EVALUATED FOR IDENTIFICATION RESULTS. ADDITIONALLY, CONTROL PANELS FROM THE SAME MATERIAL BUT DIFFERENT BATCH WERE INOCULATED WITH QC ISOLATES S. AUREUS A29213 AND S. SAPROPHYTICUS POS 432 ON A PHOENIX M50 MACHINE AND EVALUATED FOR IDENTIFICATION RESULTS. THE PANELS TESTED IDENTIFIED THEIR INOCULATED ISOLATE CORRECTLY, THEREFORE, THIS COMPLAINT IS NOT CONFIRMED FOR MISIDENTIFICATION. THE BATCH HISTORY RECORD WAS SATISFACTORY, AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. COMPLAINT TRENDING WAS PERFORMED, AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. THIS MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING APRIL 2023-2025, UNDER CAPA 11910483.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD PHOENIX¿ PID, A PATIENT SAMPLE WAS INCORRECTLY IDENTIFIED. THERE WAS NO REPORT OF PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2886997 BD PHOENIX¿ PID GRAM POSITIVE IDENTIFICATION PANEL LQL BECTON, DICKINSON & CO. (SPARKS) 4072152 30382904480081

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown