ARCHITECT ANTI-HBS
Report
- Report Number
- 3008344661-2025-00122
- Event Type
- Malfunction
- Date Received
- August 29, 2025
- Date of Event
- August 19, 2025
- Report Date
- October 22, 2025
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- LOM
- UDI-DI
- 00380740165352
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, DEVICE HISTORY RECORD REVIEW, IN HOUSE TESTING AND LABELING REVIEW. DATA AND INFORMATION PROVIDED BY THE CUSTOMER CONFIRMS THE COMPLAINT ISSUE. REVIEW OF TRACKING AND TRENDING FOR THE ARCHITECT ANTI-HBS ASSAY DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY REGARDING COMMONALITIES FOR COMPLAINT LOT AND ISSUE. ADDITIONALLY, AN INCREASE IN COMPLAINT ACTIVITY WAS NOT IDENTIFIED FOR REAGENT LOT 69530FZ01. A DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCE, POTENTIAL NON-CONFORMANCES OR DEVIATIONS ASSOCIATED WITH THE COMPLAINT LOT NUMBER AND COMPLAINT ISSUE. IN HOUSE SPECIFICITY TESTING WAS PERFORMED USING A RETAINED KIT OF THE COMPLAINT LOT. ALL ACCEPTANCE CRITERIA WAS MET AND THE PRODUCT IS PERFORMING AS EXPECTED. A REVIEW OF THE LABELLING ADDRESSES THE CUSTOMER'S ISSUE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ARCHITECT ANTI-HBS REAGENT LOT 69530FZ01 WAS IDENTIFIED.
THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, ARCHITECT ANTI-HBS, LIST NUMBER 07C18-78 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, ARCHITECT AUSAB, LIST NUMBER 01L82, WITH 510K/PMA/BLA NUMBER P050051. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT ANTI-HBS RESULT GENERATED BY THE ARCHITECT I2000SR PROCESSING MODULE FOR A PATIENT SAMPLE. UPON REPEAT TESTING, LOWER RESULTS WERE OBTAINED. THE FOLLOWING DATA WAS PROVIDED: INITIAL ARCHITECT ANTI-HBS RESULT: 12.96 MIU/ML REPEAT RESULTS: 0.86 MIU/ML AND 0.91 MIU/ML PER THE ARCHITECT ANTI-HBS PACKAGE INSERT, INTERPRETATION OF RESULTS SECTION, AN ANTI-HBS CONCENTRATION = 10 MIU/ML IS REGARDED AS BEING PROTECTIVE AGAINST HEPATITIS B VIRAL INFECTION. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT ANTI-HBS RESULT GENERATED BY THE ARCHITECT I2000SR PROCESSING MODULE FOR A PATIENT SAMPLE. UPON REPEAT TESTING, LOWER RESULTS WERE OBTAINED. THE FOLLOWING DATA WAS PROVIDED: INITIAL ARCHITECT ANTI-HBS RESULT: 12.96 MIU/ML. REPEAT RESULTS: 0.86 MIU/ML AND 0.91 MIU/ML. PER THE ARCHITECT ANTI-HBS PACKAGE INSERT, INTERPRETATION OF RESULTS SECTION, AN ANTI-HBS CONCENTRATION = 10 MIU/ML IS REGARDED AS BEING PROTECTIVE AGAINST HEPATITIS B VIRAL INFECTION. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2864545 | ARCHITECT ANTI-HBS | TEST, HEPATITIS B (B CORE, BE ANTIGEN, BE ANTIBODY, B CORE IGM) | LOM | ABBOTT IRELAND DIAGNOSTICS DIVISION | 69530FZ01 | 00380740165352 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ARC I2000SR INST, 03M74-02, (B)(6) |