FDA Adverse Event Malfunction Summary report: N

BD BBL¿ TB FLUORESCENT STAIN KIT M

MDR report key: 22916780 · Received August 28, 2025

Report

Report Number
1119779-2025-04547
Event Type
Malfunction
Date Received
August 28, 2025
Date of Event
August 1, 2025
Report Date
December 5, 2025
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
JTS
UDI-DI
00382902125198
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS MEMO IS TO SUMMARIZE FINDINGS ON YOUR COMPLAINT RELATED TO KIT TB FLUORESCENT STAIN KIT M, CATALOG NUMBER 212519, BATCH 5085995, AND COMPLAINT # (B)(4) FOR CONTAMINATION. THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF THE BATCH HISTORY RECORD TB STAIN KIT M. THE BATCH HISTORY RECORD REVIEW INDICATED NO DISCREPANCIES. ALL THE RELEASE TESTING WAS SATISFACTORY. SATISFACTORY APPEARANCE FOR AURAMINE M COMPONENT IS ¿YELLOW SUSPENSION¿. BY DEFINITION A SUSPENSION IS ¿A MIXTURE IN WHICH PARTICLES ARE DISPERSED THROUGHOUT THE BULK OF THE FLUID¿. IT IS ALSO DEFINED AS ¿A MIXTURE IN WHICH ALL PARTICLES OF A SUBSTANCE ARE DISPERSED THROUGHOUT A GAS OR LIQUID. IF A SUSPENSION IS LEFT UNDISTURBED, THE PARTICLES ARE LIKELY TO SETTLE TO THE BOTTOM. THE PARTICLES IN A SUSPENSION ARE LARGER THAN THOSE IN EITHER A COLLOID OR A SOLUTION¿. COMPLAINT TRENDS WERE REVIEWED FOR A PERIOD COVERING 12 MONTHS. DURING THAT TIME THERE HAVE BEEN NO CONFIRMED COMPLAINTS FOR THE DEFECT IN QUESTION FOR THIS PRODUCT. TWO PHOTOS AND ONE SHORT VIDEO RECEIVED IN A LEU OF RETURN. THE FIRST PHOTO DEPICTS THE FRONT OF THE BOX WITH LABEL FOR TB FLUORESCENT STAIN KIT M 212519, BATCH NUMBER AND EXPIRATION DATE NOT VISIBLE. THE SECOND PHOTO DEPICTS THE HAND HOLDING THE BOTTLE WITH YELLOW SUSPENSION AND PART OF THE LABEL FOR PART NUMBER 212514 AND EXPIRATION DATE 2025-12-31, BATCH NUMBER NOT VISIBLE. THE SHORT VIDEO SHOWS A PERSON SHAKING THE BOTTLE WITH YELLOW SUSPENSION AND PART OF THE LABEL FOR PART NUMBER 212514 AND EXPIRATION DATE 2025-12-31, BATCH NUMBER NOT VISIBLE. THE RETENTION SAMPLE FROM THE COMPLAINT BATCH NUMBER 5085995 OF TB STAIN KIT M PART NUMBER 212519, WAS EVALUATED ALONG WITH A CONTROL BATCH FOR SOLUTION APPEARANCE OF THE AURAMINE M COMPONENT PART NUMBER 212514, BATCH NUMBER 5001823 AND EXPIRATION DATE 2025-12-31. THE SOLUTION APPEARANCE OF THE RETENTION SAMPLE FOR AURAMINE M COMPONENT FROM THE KIT WAS COMPARABLE TO THE CONTROL AND WAS A YELLOW SUSPENSION. THE PRECIPITATE IN QUESTION IS INHERENT TO THE STAIN AND DOES NOT AFFECT THE PERFORMANCE OF THE STAIN. THIS COMPLAINT IS NOT CONFIRMED. BD WILL CONTINUE TO TREND DCM COMPLAINTS FOR CONTAMINATION. IF YOU HAVE FURTHER QUESTIONS, PLEASE CONTACT TECHNICAL SERVICES.

Additional Manufacturer Narrative · 0

H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED PRIOR TO USING THE BD BBL¿ TB FLUORESCENT STAIN KIT M THE USER NOTED THERE WAS WHITE SEDIMENT AT THE BOTTOM OF THE AURAMINE VIAL OF THE KIT. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED PRIOR TO USING THE BD BBL¿ TB FLUORESCENT STAIN KIT M THE USER NOTED THERE WAS WHITE SEDIMENT AT THE BOTTOM OF THE AURAMINE VIAL OF THE KIT. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2188799 BD BBL¿ TB FLUORESCENT STAIN KIT M STAINS, MICROBIOLOGIC JTS BECTON DICKINSON & CO. (SPARKS) 5085995 00382902125198

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown