COBAS® HIV-1
Report
- Report Number
- 2243471-2025-02640
- Event Type
- Malfunction
- Date Received
- August 28, 2025
- Date of Event
- May 22, 2024
- Report Date
- August 28, 2025
- Manufacturer
- ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
- Product Code
- MZF
- UDI-DI
- 00875197006407
- PMA / PMN Number
- BP150262
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE SERIAL NUMBER OF THE ANALYZER WAS (B)(6). A BATCH TREND WAS NOT IDENTIFIED FOR THE ALLEGED KIT LOT K00224. A HOLISTIC VIEW OF THE REAGENTS AND THE REAGENT KIT DOES NOT INDICATE A PRODUCT PROBLEM. BASED ON THE AVAILABLE INFORMATION, THE EXACT ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER, THE DISCREPANCY IS CONSISTENT WITH A CONTAMINATION OF THE PRIMARY TUBE AS BOTH TESTS WERE OUT OF THE SAME TUBE, WITH THE LOWER TITER BEING SECOND AFTER A FREEZE/THAW AND POTENTIAL TIME AT ROOM TEMPERATURE/MORE HANDLING. THE CUSTOMER ALSO AGREED THAT CONTAMINATION IS THE MOST LIKELY CONTRIBUTOR AS ONE OF THE SAMPLES IN THE RUN WAS AT A VERY HIGH TITER (TITER ABOVE RANGE). THE INVESTIGATION DETERMINED THAT THE DIFFERENCES OBSERVED ARE SAMPLE-SPECIFIC. INADVERTENT CONTAMINATION IS PLAUSIBLE.
THERE WAS AN ALLEGATION OF DISCREPANT COBAS® HIV-1 RESULTS FOR ONE PATIENT SAMPLE TESTED ON A COBAS 8800 ANALYZER. ON (B)(6) 2024, THE INITIAL RESULT WAS 82,000 CP/ML. THIS RESULT WAS UNEXPECTED AS THE PATIENT IS USUALLY AROUND 134 CP/ML. ON (B)(6) 2024, THE SAME SAMPLE TUBE WAS REPEATED AND THE RESULT WAS 13,600 CP/ML. DATE UNKNOWN, A NEW SAMPLE WAS OBTAINED FROM THE PATIENT AND THE RESULT WAS ACCORDING TO THE EXPECTED VALUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2520584 | COBAS® HIV-1 | HUMAN IMMUNODEFICIENCY VIRUS (HIV) VIRAL LOAD MONITORING TEST | MZF | ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG | K00224 | 00875197006407 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |