FDA Adverse Event Malfunction Summary report: N

WAVEONE GOLD PRIMARY 6-FILE STER 25MM

MDR report key: 22912012 · Received August 28, 2025

Report

Report Number
8031010-2025-00145
Event Type
Malfunction
Date Received
August 28, 2025
Date of Event
July 26, 2025
Report Date
September 16, 2025
Manufacturer
MAILLEFER INSTRUMENTS HOLDING SARL
Product Code
EKS
UDI-DI
D716WGPRIME251
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION RESULTS FOR THIS COMPLAINT ARE RETURNED WAVEONE GOLD PRIMARY FILE 25MM IS BROKEN AT THE TIP OF THE ACTIVE PART (FATIGUE). NO MATERIAL DEFECT WAS FOUND DURING ANALYSIS OF THE RUPTURE PATTERN. NO UNUSED FILE IS AVAILABLE FOR EVALUATION. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHR REVIEW (BATCH #1860947). ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND.

Additional Manufacturer Narrative · 0

AS A RESULT OF THIS MALFUNCTION, THE POTENTIAL FOR SURGICAL INTERVENTION EXISTS TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION AS EVIDENCED BY PREVIOUS REPORTED EVENTS WITH SIMILAR FILES. THIS EVENT, THEREFORE, IS REPORTABLE PER 21CFR PART 803.

Description of Event or Problem · 0

IN THIS EVENT IT IS REPORTED THAT A WAVEONE GOLD PRIMARY 6-FILE STER 25MM FILE BROKE DURING USE. THE ROOT CANAL TREATMENT WAS STOPPED DUE TO THE FILE SUDDENLY BREAKING. THE BROKEN PART WAS CAREFULLY REMOVED, ENSURING THAT NO FRAGMENTS REMAINED. FURTHER INFORMATION REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1834153 WAVEONE GOLD PRIMARY 6-FILE STER 25MM FILE, PULP CANAL, ENDODONTIC EKS MAILLEFER INSTRUMENTS HOLDING SARL 1860947 D716WGPRIME251

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown