FDA Adverse Event Injury Summary report: N

QUICK SET

MDR report key: 22910436 · Received August 28, 2025

Report

Report Number
3003442380-2025-13267
Event Type
Injury
Date Received
August 28, 2025
Date of Event
August 5, 2025
Report Date
August 30, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244017450
PMA / PMN Number
K991759
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: UNITED ARAB EMIRATES.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6001147, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 29-AUG-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL 6001147. THE COUNT OF COMPLAINTS IS 2 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6001147 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 18 AND MANUFACTURED IN THE MULTIVAC 10 ON 21-APR-2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: IN ORDER TO TEST THE PRODUCT, THE RETURNED SAMPLE(S) FROM THE LOT HAVE BEEN REQUESTED. NO PHOTO OR PHYSICAL SAMPLE WAS PROVIDED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED ARAB EMIRATES. IT WAS REPORTED THAT THE PATIENT WENT TO AN EMERGENCY ROOM (ER) AND EVENTUALLY GOT HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS (DKA) EVENT ON (B)(6) 2025. BLOOD GLUCOSE LEVEL WAS 400 MG/DL AT THE TIME OF THE EVENT AND PATIENT GOT TREATED WITH INSULIN VIA PUMP. PATIENT HAS EXPERIENCED THE SYMPTOMS OF ABDOMINAL PAIN, HEADACHE, VOMITING AT THE TIME OF THE EVENT. THE DURATION OF HOSPITALIZATION WAS LESS THAN 24 HOURS. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1519716 QUICK SET UNO QUICK-SET 60/6 SC1 MECA FPA UNOMEDICAL DEVICES S.A. DE C.V. MMT-399A 6001147 05705244017450

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention