NIPRO CELLENTIA H DIALYZER
Report
- Report Number
- 9610987-2025-00015
- Event Type
- Malfunction
- Date Received
- August 27, 2025
- Date of Event
- July 21, 2025
- Report Date
- April 9, 2026
- Manufacturer
- NIPRO CORPORATION (ODT)
- Product Code
- KDI
- UDI-DI
- 00038379008853
- PMA / PMN Number
- K160444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
INVESTIGATION REPORT ATTACHED IS ON RETAINED SAMPLES BY THE MANUFACTURER ONLY AND 2 UNUSED SAMPLES RETURNED BY THE CUSTOMER..
PENDING RESPONSE FROM THE CUSTOMER REGARDING SAMPLE AVAILABILITY, TO RETURN TO THE FACTORY FOR INVESTIGATION.
DURING HEMODIALYSIS, APPROXIMATELY 1 HOUR AND 10 MINUTES INTO TREATMENT, MACHINE ALARMED FOR BLOOD LEAK, BUT NURSE SILENCED THE ALARM SEVERAL TIMES. BLOOD LEAK TEST STRIP CONFIRMED POSITIVE FOR BLOOD LEAK. PATIENT DID NOT ENCOUNTER ANY SYMPTOMS WHEN THE LEAK OCCURRED. NO INFORMATION WAS PROVIDED ON THE PATIENT'S HISTORY. DURING TREATMENT PATIENT RECEIVED ESA (ERYTHROPOIESIS-STIMULATING AGENTS) MEDICATION. MACHINE WAS PULLED AND SWITCHED WITH A NEW ONE. DIALYZER LOT #24E0GE LINE LOT #F1YH011, PATIENT WAS ABLE TO COMPLETE THEIR PRESCRIBED TREATMENT. DIALYSIS SETTINGS ARE 4HR RUN TIME, QB AT 400, TMP= 30 TO 95, HEPARIN AT 1000U/1000U. CLINIC STATED THAT THE SAME PATIENT EXPERIENCED ANOTHER DIALYZER LEAK 23 DAYS LATER.
DURING HEMODIALYSIS, APPROXIMATELY 1 HOUR AND 10 MINUTES INTO TREATMENT, MACHINE ALARMED FOR BLOOD LEAK, BUT NURSE SILENCED THE ALARM SEVERAL TIMES. BLOOD LEAK TEST STRIP CONFIRMED POSITIVE FOR BLOOD LEAK. PATIENT DID NOT ENCOUNTER ANY SYMPTOMS WHEN THE LEAK OCCURRED. NO INFORMATION WAS PROVIDED ON THE PATIENT'S HISTORY. DURING TREATMENT PATIENT RECEIVED ESA (ERYTHROPOIESIS-STIMULATING AGENTS) MEDICATION. MACHINE WAS PULLED AND SWITCHED WITH A NEW ONE. DIALYZER LOT #24E0GE LINE LOT #F1YH011, PATIENT WAS ABLE TO COMPLETE THEIR PRESCRIBED TREATMENT. DIALYSIS SETTINGS ARE 4HR RUN TIME, QB AT 400, TMP= 30 TO 95, HEPARIN AT 1000U/1000U. CLINIC STATED THAT THE SAME PATIENT EXPERIENCED ANOTHER DIALYZER LEAK 23 DAYS LATER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1498548 | NIPRO CELLENTIA H DIALYZER | DIALYZER | KDI | NIPRO CORPORATION (ODT) | DD+CT-C-21H | 24E06E | 00038379008853 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention| O |