NIPRO CELLENTIA H DIALYZER

MDR report key: 22908080 · Received August 27, 2025

Report

Report Number
9610987-2025-00015
Event Type
Malfunction
Date Received
August 27, 2025
Date of Event
July 21, 2025
Report Date
April 9, 2026
Manufacturer
NIPRO CORPORATION (ODT)
Product Code
KDI
UDI-DI
00038379008853
PMA / PMN Number
K160444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION REPORT ATTACHED IS ON RETAINED SAMPLES BY THE MANUFACTURER ONLY AND 2 UNUSED SAMPLES RETURNED BY THE CUSTOMER..

Additional Manufacturer Narrative · 0

PENDING RESPONSE FROM THE CUSTOMER REGARDING SAMPLE AVAILABILITY, TO RETURN TO THE FACTORY FOR INVESTIGATION.

Description of Event or Problem · 0

DURING HEMODIALYSIS, APPROXIMATELY 1 HOUR AND 10 MINUTES INTO TREATMENT, MACHINE ALARMED FOR BLOOD LEAK, BUT NURSE SILENCED THE ALARM SEVERAL TIMES. BLOOD LEAK TEST STRIP CONFIRMED POSITIVE FOR BLOOD LEAK. PATIENT DID NOT ENCOUNTER ANY SYMPTOMS WHEN THE LEAK OCCURRED. NO INFORMATION WAS PROVIDED ON THE PATIENT'S HISTORY. DURING TREATMENT PATIENT RECEIVED ESA (ERYTHROPOIESIS-STIMULATING AGENTS) MEDICATION. MACHINE WAS PULLED AND SWITCHED WITH A NEW ONE. DIALYZER LOT #24E0GE LINE LOT #F1YH011, PATIENT WAS ABLE TO COMPLETE THEIR PRESCRIBED TREATMENT. DIALYSIS SETTINGS ARE 4HR RUN TIME, QB AT 400, TMP= 30 TO 95, HEPARIN AT 1000U/1000U. CLINIC STATED THAT THE SAME PATIENT EXPERIENCED ANOTHER DIALYZER LEAK 23 DAYS LATER.

Description of Event or Problem · 0

DURING HEMODIALYSIS, APPROXIMATELY 1 HOUR AND 10 MINUTES INTO TREATMENT, MACHINE ALARMED FOR BLOOD LEAK, BUT NURSE SILENCED THE ALARM SEVERAL TIMES. BLOOD LEAK TEST STRIP CONFIRMED POSITIVE FOR BLOOD LEAK. PATIENT DID NOT ENCOUNTER ANY SYMPTOMS WHEN THE LEAK OCCURRED. NO INFORMATION WAS PROVIDED ON THE PATIENT'S HISTORY. DURING TREATMENT PATIENT RECEIVED ESA (ERYTHROPOIESIS-STIMULATING AGENTS) MEDICATION. MACHINE WAS PULLED AND SWITCHED WITH A NEW ONE. DIALYZER LOT #24E0GE LINE LOT #F1YH011, PATIENT WAS ABLE TO COMPLETE THEIR PRESCRIBED TREATMENT. DIALYSIS SETTINGS ARE 4HR RUN TIME, QB AT 400, TMP= 30 TO 95, HEPARIN AT 1000U/1000U. CLINIC STATED THAT THE SAME PATIENT EXPERIENCED ANOTHER DIALYZER LEAK 23 DAYS LATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1498548 NIPRO CELLENTIA H DIALYZER DIALYZER KDI NIPRO CORPORATION (ODT) DD+CT-C-21H 24E06E 00038379008853

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention| O