FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (HS-TNL) CARTRIDGE

MDR report key: 22904007 · Received August 27, 2025

Report

Report Number
2245578-2025-00101
Event Type
Malfunction
Date Received
August 27, 2025
Date of Event
August 17, 2025
Report Date
September 30, 2025
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
MMI
PMA / PMN Number
K240984
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 26-SEP-2025. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AO (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED FOR HS-TNI CARTRIDGE LOT A25126A.

Description of Event or Problem · 0

ON (B)(6) 2025, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT FALSE POSITIVE RESULTS ON A 36-YEAR-OLD FEMALE PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION, NOR DETAILS SURROUNDING THE EVENT AT THE TIME OF THIS REPORT. METHOD: I-STAT HS-TNL, DATE: (B)(6) 2025, COLLECTED: 1742, RESULT: 1742, SAMPLE: 545.1 NG/L WB LITHIUM HEPARIN, POSITIVE/NEGATIVE: POSITIVE. METHOD: BECKMAN HS-TNL, DATE: (B)(6) 2025, COLLECTED: 17:21, RESULT: 18:12, SAMPLE: <2.3 NG/L, POSITIVE/NEGATIVE: NEGATIVE. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THERE IS NO EVIDENCE THE PATIENT SUFFERED AN MI. PER I-STAT SYSTEM MANUAL: ART: 715595-00V, REPORTABLE RANGE: 0.00 - 50.00, REFERENCE RANGE: 0.00 - 0.03 (REPRESENTS THE 0 TO 97.5% RANGE OF RESULTS), REFERENCE RANGE: 0.00 - 0.08 (REPRESENTS THE 0 TO 99% RANGE OF RESSULTS).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1251272 I-STAT TROPONIN (HS-TNL) CARTRIDGE HS-TNL CARTRIDGE MMI ABBOTT POINT OF CARE INC. NA A25126A

Patients

Seq Age Sex Outcome Treatment
1 36 YR Female