FDA Adverse Event Malfunction Summary report: N

POWERPICC SOLO2, 3CG

MDR report key: 22900146 · Received August 26, 2025

Report

Report Number
3006260740-2025-06272
Event Type
Malfunction
Date Received
August 26, 2025
Date of Event
August 11, 2025
Report Date
October 21, 2025
Manufacturer
C.R. BARD, INC. (BASD) -3006260740
Product Code
LJS
UDI-DI
00801741138973
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. EACH EVENT REPORTED TO BD IS EVALUATED AND INVESTIGATED IN ACCORDANCE WITH OUR COMPLAINT INVESTIGATION PROCEDURES. THE INVESTIGATION PROCESS INCLUDES, BUT IS NOT LIMITED TO, EVALUATION OF THE EVENT DETAILS PROVIDED BY THE COMPLAINT FACILITY, A REVIEW OF COMPLAINT HISTORY, MANUFACTURING RECORDS AND RISK DOCUMENT WHERE APPLICABLE, AND AN EVALUATION OF THE SUBJECT DEVICE WHEN AVAILABLE TO IDENTIFY POTENTIAL CONTRIBUTING FACTORS. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION.

Additional Manufacturer Narrative · 0

H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. THE COMPLAINT OF A LEAK WAS CONFIRMED. THE PRODUCT RETURNED FOR EVALUATION WAS ONE 4 FR SINGLE LUMEN POWERPICC SOLO. THE INVESTIGATION FINDINGS ARE CONSISTENT WITH DAMAGE ACCUMULATED THROUGH FLEXURAL FATIGUE. FLEXURAL FATIGUE MAY OCCUR DUE TO CYCLIC KINKING OF THE CATHETER TUBE IN WHICH PHYSIOLOGICAL, PLACEMENT, USAGE, AND MECHANICAL FACTORS MAY GRADUALLY FORM A CRACK(S) IN THE CATHETER. THE RETURNED PRODUCT SAMPLE WAS EVALUATED, AND A SPLIT WAS OBSERVED AT THE 2 CM DEPTH MARKERS. THE CATHETER SPLIT CONTAINED PHYSICAL FEATURES ASSOCIATED WITH MATERIAL FATIGUE, AND THE CHARACTERISTICS OBSERVED WHICH SUPPORTED THIS TYPE OF FAILURE INCLUDED: ¿ DAMAGE WHICH WAS CIRCUMFERENTIALLY ALIGNED ¿ FRACTURE EDGES WHICH WERE ROUNDED AND POLISHED DUE TO REPEATED MATERIAL WEAR ¿ 'C' SHAPED IMPRESSIONS LEADING INTO THE FRACTURE SITE WHICH ARE CONSISTENT WITH MATERIAL BUCKLING DUE TO MOVEMENT WHICH CAUSED THE FRACTURE EDGES TO BE PRESSED TOGETHER ¿ OVERALL ELLIPTICAL SHAPE TO THE FRACTURE CROSS-SECTION (A RESULT OF REPEATED KINKING OF THE TUBING) A FLATTENED SECTION OF CATHETER WAS FOUND BETWEEN 2.5 CM AND 4.5 CM. REPETITIVE KINKING OF THE INDWELLING CATHETER APPEARS TO HAVE CONTRIBUTED TO THE OBSERVED LEAK; HOWEVER, THE SPECIFIC CIRCUMSTANCES SURROUNDING HOW KINKING DEVELOPED INTO A FRACTURE WITHIN THE CATHETER TUBING WERE ULTIMATELY UNKNOWN. THIS COMPLAINT WILL BE RECORDED FOR FUTURE TRENDING AND MONITORING PURPOSES.

Description of Event or Problem · 0

IT WAS REPORTED PATIENT ATTENDS TODAY FOR ASSESSMENT OF HIS PICC AS DN UNABLE TO GET A GOOD VENOUS RETURN. WHEN ASSESSED BY SN UNABLE TO FLUSH & NO VENOUS RETURN- SLIGHT BUBBLING NOTICED & DRIED BLOOD IN THE LINE. FRESH BLOOD NOTED UNDER DRESSING ALSO. DN HAD REDRESSED LINE TODAY. ASKED PICC PLACER FOR ADVICE & SHE CAME & ASSESSED PICC. PICC LINE FRACTURED AT MARK 2. LINE REMOVED MEASURING AT 45CM. NO ISSUES WITH REMOVAL. FOR REINSERTION SAME DAY AT 1430. PICC INSERTED TO BASILIC VEIN (B)(6) 2025, EXIT MARKING 45CM. NO HISTORY OF OCCLUSIONS, UNSURE IF USED FOR CT CONTRAST. PICC REMOVED (B)(6) 2025, PATIENT ON CAPOX, NOT RUNNING ON DAY OF FRACTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2780112 POWERPICC SOLO2, 3CG PERIPHERALLY INSERTED CENTRAL CATHETER LJS C.R. BARD, INC. (BASD) -3006260740 N/A REJT0680 00801741138973

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other