FDA Adverse Event Malfunction Summary report: N

DA VINCI XI

MDR report key: 22899278 · Received August 26, 2025

Report

Report Number
2955842-2025-35577
Event Type
Malfunction
Date Received
August 26, 2025
Date of Event
August 6, 2025
Report Date
November 6, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874115343
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) RECEIVED A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. SEVERAL ERBE ERRORS (M-26, M-11, M-10, M-12, M-02) WERE CONFIRMED IN THE LOGS. THE ERBE WAS TESTED USING THE SYSTEM, AND ERROR M-B0-60 SHOWED UP WHEN TESTING MONOPOLAR. THE COMPLAINT WAS CONFIRMED BASED ON THE FIELD EVALUATION AND FAILURE ANALYSIS. THE PROBABLE ROOT CAUSE OF ERROR M-36 IS ATTRIBUTED TO INSTALLATION ISSUES PREVENTING ENERGY ACTIVATION, AND M-11 INDICATES A CPU AND SENSORS MODULE INTERNAL TIME OUT. THE INSTRUMENT MAY BE INSTALLED ON THE PATIENT SIDE CART ARM, BUT THE ENERGY CORD MAY NOT BE PROPERLY CONNECTED TO THE INSTRUMENT AND/OR GENERATOR. THIS ERROR CAN BE RESOLVED THROUGH TROUBLESHOOTING OR REPLACEMENT OF THE GENERATOR.

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE REPLACED THE INTEGRATED ELECTROSURGICAL UNIT (IESU) TO RESOLVE THE ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID NOT YET RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE COMPLAINT WAS CONFIRMED BASED ON THE FIELD EVALUATION. THE PROBABLE ROOT CAUSE IS ATTRIBUTED TO A DEFECTIVE IESU. THE FSE REPLACED THE PART TO RESOLVE THE ISSUE.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE ERBE WAS NOT WORKING. THE CUSTOMER WAS GUIDED TO INSTALL AN EXTERNAL GENERATOR TO PROCEED. THE PROCEDURE WAS COMPLETING AS PLANNED WITH NO REPORTED INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1406554 DA VINCI XI VISION SIDE SYSTEM, RECONDITIONED NAY INTUITIVE SURGICAL, INC 380721-09 N/A 00886874115343

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES.